![Transcell Technology TI-500 Plus Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/transcell-technology/ti-500-plus/ti-500-plus_setup-and-operation-manual_1150672069.webp)
7
CODE/NAME
DESCRIPTION
SELECTION LIST
F5
Motion Band
Sets the level at which motion is detected. If motion is not detected,
the scale can process a Print or Zero command. Maximum value
varies depending on local regulations. Expressed as scale divisions
per second (d/s).
1d
√
√√
√
3d
5d
10d
F6
Digital Filter
Averages weight readings to produce higher stability. Choose the
speed that works best for your application.
“FAST” = Fast “nnEd” = Medium “SLo” = Slow
FAST
nnEd
√
√√
√
SLo
F7
Overload Limit
Selects the desired formula which determines the point at which the
indicator shows overload. All selections are based on the primary
unit selected in F8.
"FS" = Full scale capacity.
FS
FS + 2%
√
√√
√
FS + 1d
FS + 9d
F8
Calib. Unit
Selects the primary base unit to be used in the calibration process.
Also the default unit for normal operation.
"1" = primary unit is lb. "2" = primary unit is in kg.
1
√
√√
√
2
F9
Display
Divisions
Determines the desired weight increments. Value should be
consistent with legal requirements.
1
√
√√
√
2
5
F10
Decimal Pt.
Determines location of the decimal point.
0
√
√√
√
0.0
0.00 0.000
0.0000 00
F16
Zero
Calibration
Places indicator into the zero calibration routine. Scrolling down with
the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F17
Span
Calibration
Places indicator into the span calibration routine. Scrolling down
with the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F18
View
Calibration
Actuates the function that allows you to view both the zero and span
calibration value. The values displayed in this function are valid only
after Calibration (F16 & F17) has been successfully completed.
Scrolling down with the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F19
Key-in Zero
Allows you to key-in known zero calibration value in case of memory
loss in the field. Scrolling down with the
ZERO
key one level begins
the procedure.
Press
ZERO
key
to begin sequence
F20
Key-in Span
Allows you to key-in a known span calibration value in case of
memory loss in the field. Scrolling down with the
ZERO
key one
level begins the procedure.
Press
ZERO
key
to begin sequence
F21
Factory Reset
This sub-menu will reset all parameters in the “F” and “A” menu to
the default settings. USE WITH CAUTION!
Press the
ZERO
key twice to
execute
.
F24
Piece Count
Enable
Used to enable or disable the piece count feature. This feature must
be disabled for commercial applications.
"0" = Disable piece count "1" = Enable piece count
0
√
√√
√
1
Содержание TI-500 Plus
Страница 22: ...i TI 500E Plus Digital Weight Indicator Setup Operation Manual Revision 1 2 September 16 2011...
Страница 44: ...i Digital Weight Indicator Setup OperationManual Revision 1 1 April 6 2009...
Страница 62: ...TI 500E Digital Weight Indicator Setup Operation Manual Revision 3 0 July 6 2012...