11
5.
Repeat steps 2 thru 4 for C2 and C3. At the conclusion of C3, the indicator reverts back up to
F17. NOTE: If you wish to use only one calibration point (C1), simply press the NET/GROSS
key when prompted for C2 and C3 (do not enter in a calibration value).
6.
At this time it is suggested that the calibration values be recorded for future use (see next
section).
If the calibration was
not
successful, one of the error messages below will appear. Take the indicated
action to correct the problem, then perform a new calibration.
"
Err0
" - The calibration test weight or the keyed-in weight is larger than the full capacity of the
scale. Change the calibration test weight or check the input data.
"
Err1
" - The calibration test weight or the keyed-in weight is smaller than 1% of the full capacity
of the scale. Change the calibration test weight or check the input data.
"
Err2
" – There is not enough signal from the load cells to establish a proper calibration. Most
commons causes include incorrect load cell wiring, a mechanical obstruction or a faulty load
cell.
VIEW CALIBRATION VALUES (F18)
Note:
The values displayed in this procedure are valid only after a successful calibration has been
performed using F16 and F17.
1.
While in the Setup mode, scroll to "
F 18
", then scroll down once using the ZERO key to enter
View calibration menu.
2.
The display will show the information listed in Table 2. The code will display briefly followed by
the value. It is recommended that you record each value in the table below. Press any key to
continue down the list. At the completion of the list, the indicator reverts back up to F18.
CODE
NAME
VALUE
C 0
Zero Calibration Value
T 1
First Test Weight Value
C 1
First Span Calibration Value
T 2
Second Test Weight Value
C 2
Second Span Calibration Value
T 3
Third Test Weight Value
C 3
Third Span Calibration Value