CONNECTING THE WEIGH PLATFORM
The TI-500 ships with a 15 ft shielded load cell cable for connection to weigh platform’s load cell(s) or
junction box.
1. Plug the cable’s 14-pin parallel interface connector into the load cell port on the rear
panel of the indicator.
2. Wire the bare wires and shield to the weigh platform’s load cell(s) or junction box using
the color codes shown below.
4-wire cable
Optional 6-wire cable
Color
Wire Name
Color
Wire Name
Red
+ Excitation
Red
+Excitation
Black
- Excitation
Black
-Excitation
Green
+ Signal
Green
+Signal
White
- Signal
Yellow
- Signal
Orange
+ Sense
Brown
- Sense
CONNECTING THE SERIAL I/O DEVICE
The TI-500 model comes standard with one full duplex RS-232 serial port, designed for connection to a
computer or a serial printer. The same port may be also used as a simplex, RS-232 port designed for
connection to a remote display.
DSUB9 Connector
Pin No.
Wire Name
2
RXD
3
TXD
5
Ground
CONNECTING THE POWER SUPPLY
The TI-500 indicator ships standard with an external AC adapter.
1. Simply plug the AC adapter into the indicator’s DC Power Jack first, and then plug
into a standard wall outlet. Make sure that the AC voltage appearing at the wall
outlet matches the input voltage marked on the AC adapter.