APPENDIX C: DISPLAYED ERROR CODES 

 

CODE 

MODE 

MEANING / POSSIBLE SOLUTION 

 

Normal Operating 
Mode 

Gross  Overload.  A  weight  greater  than  the  rated  capacity  has  been 
applied  to  the  scale.  Remove  the  weight  from  the  platter  or  try  re-
calibrating the scale. Otherwise, check for a bad load cell connection or 
possible load cell damage due to overloading.

 

Err 0 

Span Calibration 
Mode (F17) 

Keyed-in  weight  value  is  larger  than  full-scale  capacity.  Use  a  smaller 
test weight or check keyed-in value. 

Err 1 

Span Calibration 
Mode (F17) 

Keyed-in weight value is less than 1% of full-scale capacity. Use a larger 
test weight or check keyed-in value. 

Err 2 

Span Calibration 
Mode (F17) 

There  is  not  enough  load  cell  signal  to  produce  the  internal  counts 
necessary  to  properly  calibrate  the  scale.  First  check  all  load 
connections. Use F16 mode to view internal counts.  

Err 3 

All Modes 

Non-volatile  memory  read  error.  One  or  more  setup  parameters  have 
been lost. 

Err 4 

All Modes 

Non-volatile memory write error. Indicator needs service. 

Err 5 

Key-in Span 
Calibration Mode 
(F20) 

You  have  attempted  to  enter  a  zero  value  for  C1-C3.  Enter  a  known 
calibration value greater than zero. 

Err 9 

Normal Operating 
Mode 

Span calibration value has been lost. Re-calibrate the scale. 

 

 

Содержание TI-500 Plus

Страница 1: ...i TI 500 Plus DIGITAL WEIGHT INDICATOR Setup Operation Manual Revision 1 0 975 Deerfield Parkway Buffalo Grove IL 60089 April 3 2008 Tel 847 419 9180 Fax 847 419 1515 http www transcell net...

Страница 2: ...ITEM 14 PIECE COUNTING MODE 14 PEAK HOLD MODE 15 REMOTE DISPLAY MODE 15 LEGAL FOR TRADE SEALING 15 APPENDIX A SPECIFICATIONS 16 APPENDIX B SERIAL PORT INFORMATION 17 SERIAL PORT MODES 17 DEMAND DUPLE...

Страница 3: ...ce when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions man...

Страница 4: ...er must first be removed to make the appropriate connections to the weigh platform printer remote display and power supply To remove the rear cover simply remove the screws that secure it to the enclo...

Страница 5: ...ex RS 232 port designed for connection to a remote display DSUB9 Connector Pin No Wire Name Pin No Wire Name 2 RXD 7 RTS 3 TXD 8 CTS 5 Ground Connection assignments for all serial RS 232 communication...

Страница 6: ...r off the indicator 2 Press and hold the SET key for about three seconds while turning the unit back on The display shows F 1 to indicate that you are in Setup Menu mode 3 Use the navigation keys show...

Страница 7: ...and environmental limits on the useful system resolution Pressing the ZERO key to scroll down one level begins the sequence Key in 100 50000 5000 NA 3000 F2 Sampling Rate Sets the sampling rate in He...

Страница 8: ...es indicator into the span calibration routine Scrolling down with the ZERO key one level begins the procedure Press ZERO key to begin sequence F18 View Calibration Actuates the function that allows y...

Страница 9: ...25d US 0 5d F34 Auto Print Min Weight Selects the minimum weight at which the auto print function will work if enabled Selections are in display divisions d Scrolling down with the ZERO key one level...

Страница 10: ...l digit segments decimal points and LCD annunciators in a test sequence Press ZERO key to begin sequence A6 Output String Selects fixed output string for serial port Standard Chipset only Refer to App...

Страница 11: ...ease the flashing digit by pressing the UNITS key Decrease the flashing digit by pressing the ZERO key Pressing the PRINT key or the TARE key will change the position of the flashing digit USER MODE K...

Страница 12: ...key to enter zero calibration menu The display will momentarily show C 0 followed by a value This value is the internal A D count and can prove useful when trying to troubleshoot setup problems 2 Afte...

Страница 13: ...or check the input data Err2 There is not enough signal from the load cells to establish a proper calibration Most commons causes include incorrect load cell wiring a mechanical obstruction or a fault...

Страница 14: ...span calibration values obtained from a successful F17 calibration procedure must be used 1 While in the Setup mode scroll to F 20 and then scroll down once using the ZERO key The indicator will prom...

Страница 15: ...dicator is displaying net weight G Indicates that the indicator is displaying gross weight T Indicates that a tare weight has been established in the system lb kg Indicates the unit of the displayed w...

Страница 16: ...If necessary press the ZERO key to obtain a weight reading of zero 3 If weighing an item in a container place the empty container on the scale s platter and after allowing the weight indication to sta...

Страница 17: ...for your scale to process accurately 5 To exit the piece count mode press the UNITS key NOTE The APW will NOT remain in scale memory when you exit piece counting mode PEAK HOLD MODE IMPORTANT NOTE Th...

Страница 18: ...hipset EEPROM 256 x 8 external to C Digital Filtering Software selectable SERIAL COMMUNICATIONS Serial Port Full Duplex selectable Baud rate 8 data bits no parity 1 stop bit 7 data bits odd parity 1 s...

Страница 19: ...l not respond if the scale is in motion positive overload or negative overload Z This command is sent to the indicator to zero the scale The indicator will not respond if the scale is in motion positi...

Страница 20: ...or two way communication STX POL xxxxx xx LB KG GR NT CR LF Polarity SP Positive Negative Weight Data Units LB pound KG kilogram Gross Net GR Gross NT Net Start Transmission Carriage Return Line Feed...

Страница 21: ...Keyed in weight value is less than 1 of full scale capacity Use a larger test weight or check keyed in value Err 2 Span Calibration Mode F17 There is not enough load cell signal to produce the intern...

Страница 22: ...i TI 500E Plus Digital Weight Indicator Setup Operation Manual Revision 1 2 September 16 2011...

Страница 23: ...TION F17 10 VIEW CALIBRATION VALUES F18 11 KEY IN ZERO CALIBRATION VALUE F19 12 KEY IN SPAN CALIBRATION VALUE F20 12 OPERATION 13 DISPLAY 13 KEYBOARD 14 FUNCTION KEYS 14 GENERAL SCALE OPERATION 14 WEI...

Страница 24: ...ial installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communic...

Страница 25: ...oad cell connections ground shield wire at indicator and run these cables away from your AC DC power cables if possible In extreme cases it may be necessary to install surge suppressors line condition...

Страница 26: ...een Signal White Signal Yellow Signal Orange Sense Brown Sense CONNECTING THE SERIAL I O DEVICE The TI 500E Plus model comes standard with one full duplex RS 232 serial port designed for connection to...

Страница 27: ...over and move it to the opposite position NOTE A metal plate held on by two drilled head screws may conceal the slide switch 3 Power on the indicator The display shows F 1 to indicate that you are in...

Страница 28: ...100 50000 5000 NA 3000 F2 Sampling Rate Sets the sampling rate in Hertz measurements per second Use 10 Hz for most applications or 80 Hz for extra fast response time 10 NA 80 F3 Zero Track Band Selec...

Страница 29: ...fter Calibration F16 F17 has been successfully completed Scrolling down with the ZERO key one level begins the procedure Multi point cal Press ZERO key to begin sequence F19 Key in Zero Allows you to...

Страница 30: ...level begins the procedure 0 Disabled Key in 0 100 1 NA SETUP MENU PROCEDURES This section provides instructions for all of the Setup Menu procedures except F16 through F20 whose procedures can be fou...

Страница 31: ...t segments decimal points and LCD annunciators in a test sequence Press ZERO key to begin sequence A6 Output String Selects fixed output string for serial port Standard Chipset only Refer to Appendix...

Страница 32: ...e display will show SET momentarily and then revert back up to A8 LF Line Feeds Number Entry A9 1 While in the User Menu mode scroll to A 9 and then scroll down once using the ZERO key to enter the Li...

Страница 33: ...troubleshoot setup problems 2 After making sure that there are no test weights on the platform press the ZERO key again to zero out the displayed value 3 Press the NET GROSS key to save the zero point...

Страница 34: ...ons causes include incorrect load cell wiring a mechanical obstruction or a faulty load cell VIEW CALIBRATION VALUES F18 Note The values displayed in this procedure are valid only after a successful c...

Страница 35: ...id span calibration values obtained from a successful F17 calibration procedure must be used 1 While in the Setup mode scroll to F 20 and then scroll down once using the ZERO key The indicator will pr...

Страница 36: ...ght is active whenever the displayed weight is within a pre programmed band from true zero NET Indicates that the indicator is displaying net weight GROSS Indicates that the indicator is displaying gr...

Страница 37: ...in motion 3 The scale is not in overload see Appendix D for error codes Print This key is used to send weight information out to the serial port provided the following conditions are met 1 The scale...

Страница 38: ...ing mode and you must start again from Step 2 3 Place the sample items on the platform all at once and allow the weight indication to stabilize Once this is done the zero indicated after the sample nu...

Страница 39: ...ously and at the same transmission baud rate configured in A1 LEGAL FOR TRADE SEALING Indicators can be sealed for commercial Legal for Trade applications as follows 1 Power off the indicator 2 On the...

Страница 40: ...ard Chipset 1024 x 8 internal to C Enhanced Chipset EEPROM 256 x 8 external to C Digital Filtering Software selectable SERIAL COMMUNICATIONS Serial Port Full Duplex selectable Baud rate 8 data bits no...

Страница 41: ...tor to zero the scale The indicator will not respond if the scale is in motion positive overload or negative overload The indicator will also not respond if it is not in gross mode or within the zero...

Страница 42: ...or two way communication STX POL xxxxx xx LB KG GR NT CR LF Polarity SP Positive Negative Weight Data Units LB pound KG kilogram Gross Net GR Gross NT Net Start Transmission Carriage Return Line Feed...

Страница 43: ...Keyed in weight value is less than 1 of full scale capacity Use a larger test weight or check keyed in value Err 2 Span Calibration Mode F17 There is not enough load cell signal to produce the intern...

Страница 44: ...i Digital Weight Indicator Setup OperationManual Revision 1 1 April 6 2009...

Страница 45: ...ONS 13 APPENDIX B SERIAL PORT INFORMATION 14 SERIAL PORT MODES 14 DEMAND DUPLEX MODE 14 CONTINUOUS DUPLEX MODE 14 OUTPUT STRINGS 15 TEXT PRINT TICKET 15 STRING FORMAT 1 Condec Demand String 15 STRING...

Страница 46: ...so it will be available if you have questions concerning the operation of the scale MODEL DISPLAY TYPE ENCLOSURE TYPE POWER SOURCE TI 500 LCD Mild Steel ABS AC adapter 6 VDC 300 mA or 4 AA batteries...

Страница 47: ...oad cell connections ground shield wire at indicator and run these cables away from your AC DC power cables if possible In extreme cases it may be necessary to install surge suppressors line condition...

Страница 48: ...on Green Signal Green Signal White Signal Yellow Signal Orange Sense Brown Sense CONNECTING THE SERIAL I O DEVICE The TI 500 model comes standard with one full duplex RS 232 serial port designed for c...

Страница 49: ...the front panel keys become directional navigators to move around in the menus and one key is used to save or SET the selections ACCESSING THE MENUS To access the Setup F menu 1 Power off the indicat...

Страница 50: ...ving press the UNITS up key to return to the current heading 5 Repeat Steps 2 through 5 until the Menu is programmed SETUP MENU DESCRIPTIONS This section provides more detailed descriptions of the sel...

Страница 51: ...requirements 1 2 5 F10 Decimal Pt Determines location of the decimal point 0 0 0 0 00 0 000 0 0000 00 F16 Zero Calibration Places indicator into the zero calibration routine Scrolling down with the ZE...

Страница 52: ...rs in a test sequence Pressing the ZERO key to scroll down one level begins the test sequence Press ZERO key to begin sequence A5 Disable the lb kg Key Allows the lb kg key to be disabled so that an o...

Страница 53: ...O key Pressing the PRINT key or the TARE key will change the position of the flashing digit 4 After setting the exact value press the NET GROSS key to save the ID Number value The display will show SE...

Страница 54: ...a value This value is the internal A D count and can prove useful when trying to troubleshoot setup problems 2 After making sure that there are no test weights on the platform press the ZERO key again...

Страница 55: ...he list the indicator reverts back up to F18 ZERO CALIBRATION VALUE C0 SPAN CALIBRATION VALUE C1 Table 2 Calibration Value Table KEY IN ZERO CALIBRATION VALUE F19 Note This procedure is intended for e...

Страница 56: ...ght is within 0 25 divisions of true zero N Indicates that the indicator is displaying net weight G Indicates that the indicator is displaying gross weight T Indicates that a tare weight has been esta...

Страница 57: ...scale is not in overload see Appendix D for error codes On Print This key is used switch the indicator on it can also send weight information out to the serial port provided the following conditions a...

Страница 58: ...of FS Calibration Method Software Calibration with long term storage in EEPROM Excitation Voltage 3 3 VDC 4 x 350 load cells DIGITAL SPECIFICATIONS Microcontroller PIC Program Memory 32K x 8 internal...

Страница 59: ...sent to the indicator to print the indicated display The indicator will not respond if the scale is in motion positive overload or negative overload Z This command is sent to the indicator to zero th...

Страница 60: ...CR LF Polarity SP Positive Negative Weight Data Units LB pound KG kilogram Gross Net GR Gross NT Net Start Transmission Carriage Return Line Feed SP SP Space Space pc pieces cu cusotm unit pcs pieces...

Страница 61: ...l to produce the internal counts necessary to properly calibrate the scale First check all load connections Use F16 mode to view internal counts Err 3 All Modes Non volatile memory read error One or m...

Страница 62: ...TI 500E Digital Weight Indicator Setup Operation Manual Revision 3 0 July 6 2012...

Страница 63: ...VERVIEW 10 ZERO CALIBRATION F16 10 SPAN CALIBRATION F17 10 VIEW CALIBRATION VALUES F18 11 KEY IN ZERO CALIBRATION VALUE F19 11 KEY IN SPAN CALIBRATION VALUE F20 11 OPERATION 13 DISPLAY 13 KEYBOARD 14...

Страница 64: ...ial installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communic...

Страница 65: ...ections ground shield wire at indicator and run these cables away from your AC DC power cables if possible In extreme cases it may be necessary to install surge suppressors line conditioners or even U...

Страница 66: ...ation Green Signal Green Signal White Signal Yellow Signal Orange Sense Brown Sense CONNECTING THE SERIAL I O DEVICE The TI 500E model comes standard with one full duplex RS 232 serial port designed f...

Страница 67: ...tors to move around in the menus and one key is used to save or SET the selections ACCESSING THE MENUS To access the Setup F menu 1 Power off the indicator 2 Locate the slide switch on the rear cover...

Страница 68: ...urrent heading 5 Repeat Steps 2 through 5 until the Menu is programmed SETUP MENU DESCRIPTIONS This section provides more detailed descriptions of the selections found in the Setup Menu Chart Factory...

Страница 69: ...rolling down with the ZERO key one level begins the procedure Press ZERO key to begin sequence F17 Span Calibration Places indicator into the span calibration routine Scrolling down with the ZERO key...

Страница 70: ...O key to begin sequence A5 Disable the Units Key Allows the Units key to be disabled so that an operator cannot accidentally press the key and change the displayed units lb kg 0 Disable the Units key...

Страница 71: ...mentarily and then revert back up to A8 LF Line Feeds Number Entry A9 1 While in the User Menu mode scroll to A 9 and then scroll down once using the ZERO key to enter the Line Feeds menu 2 The displa...

Страница 72: ...lowed by a value This value is the internal A D count and can prove useful when trying to troubleshoot setup problems 2 After making sure that there are no test weights on the platform press the ZERO...

Страница 73: ...any key to continue down the list At the completion of the list the indicator reverts back up to F18 ZERO CALIBRATION VALUE C0 SPAN CALIBRATION VALUE C1 Table 2 Calibration Value Table KEY IN ZERO CA...

Страница 74: ...12 4 If the entered values are entered successfully the display will show E CAL 1 momentarily before reverting back up to F20...

Страница 75: ...ever the displayed weight is within a pre programmed band from true zero NET Indicates that the indicator is displaying net weight GROSS Indicates that the indicator is displaying gross weight TARE In...

Страница 76: ...in motion 3 The scale is not in overload see Appendix D for error codes Print This key is used to send weight information out to the serial port provided the following conditions are met 1 The scale...

Страница 77: ...indicated on the display If the screen does not show 5 0 press the ZERO key once The indicator is prompting you to place five identical items on the platform NOTE If you wish to change the sample numb...

Страница 78: ...plications as follows 1 Power off the indicator 2 On the back of the indicator locate the setup calibration switch cover 3 Thread a wire security seal through both drilled head screws securing the cal...

Страница 79: ...8 internal to C EEPROM 126 x 8 external to C Digital Filtering Software selectable SERIAL COMMUNICATIONS Serial Port Full Duplex selectable Baud rate 8 data bits no parity 1 stop bit 7 data bits odd p...

Страница 80: ...sent to the indicator to print the indicated display The indicator will not respond if the scale is in motion positive overload or negative overload Z This command is sent to the indicator to zero th...

Страница 81: ...s as a CTS pin Refer to the printer s user manual to confirm which pin is the DTR pin NOTES 1 The TARE and NET fields are not printed unless a tare has been established in the system 2 The ID number f...

Страница 82: ...damage due to overloading Err 0 Span Calibration Mode F17 Keyed in weight value is larger than full scale capacity Use a smaller test weight or check keyed in value Err 1 Span Calibration Mode F17 Ke...

Отзывы: