TH2883 Series Operation Manual Chapter 5 Introduction to [SETUP]
5-3
the cursor will move up, down, left and right.
5.2.3 CH1-CH8(TH2883S8-5), CH1-CH4(TH2883S4-5) Channel
setup
TH2883S8-5 provides 8 channels from CH1 to CH8, TH2883S4-5 provides 4 channels from
CH1 to CH4. Move the cursor to the corresponding channel, CLOSE, HIGH and LOW can be set
for this channel. Red means the current mode of the channel is HIGH. Black means the current
mode of the channel is LOW. White means the current mode of the channel is CLOSE.
CLOSE Close this test channel.
HIGH Set the mode of this channel as HIGH.
LOW Set the mode of this channel as LOW.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6,
the cursor will move up, down, left and right.
5.2.4 Samp- can be set in TEST mode
200MSa/s, 100MSa/s, 50MSa/s, 25MSa/s, 12.5MSa/s, 6.25MSa/s, 3.12MSa/s, 1.56MSa/s are
optional for sample rate. Move the cursor to the sample rate zone, following soft keys are
available.
Be used to select the desired sample rate.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the
cursor will move up, down, left and right.
5.2.5 Test Imp- can be set in TEST mode
The number of tested impulses can be set from 1 to 32. Move the cursor to this zone, the
following soft keys will be displayed.
They are fine adjustment soft keys used to change the number of impulses by 1.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the
cursor will move up, down, left and right.
Note: User can directly input the desired number by numerical keys. In the state of
inputting numbers, soft keys are unavailable. ENTER can be used to terminate
inputting.
5.2.6 Erase Imp- can be set in TEST mode
The number of erase impulses can be set from 0 to 16. Move the cursor to this zone, following
soft keys will be displayed.