TH2883 Series Operation Manual Chapter 5 Introduction to [SETUP]
5-4
They are fine adjustment soft keys used to change the number of impulses by 1.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the
cursor will move up, down, left and right.
Note: User can directly input the desired number by numerical keys. In the state of
inputting numbers, soft keys are unavailable. ENTER can be used to terminate inputting.
5.2.7 Volt ADJ- can be set in TEST mode
For different load on each DUT, when a given impulse voltage is applied across them, the actual
voltage applied on each DUT may also be different. User can turn on the voltage adjust function
to automatically change the output voltage, thus making all voltages applied across all DUT
identical.
Move the cursor to this zone, following soft keys will be available.
ON
Turn on the voltage adjust function.
OFF
Turn off the voltage adjust function.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the
cursor will move up, down, left and right.
5.2.8 Start Volt- can be set in BDV test mode
Start Volt ranges from 100V to 5000V but cannot be larger than End Volt. Move the cursor to this
zone, following soft keys will be displayed.
They are coarse adjustment keys used to change Start Volt by 100V.
They are fine adjustment keys used change Start Volt by 10V.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the
cursor will move up, down, left and right.
5.2.9 End Volt- can be set in BDV test mode
End Volt ranges from 100V to 5000V but must be larger than Start Volt. Move the cursor to this
zone, following soft keys will be displayed.
They are coarse adjustment keys used to change End Volt by 100V.
They are fine adjustment keys used change End Volt by 10V.
Lighten this key to enable the numerical arrow key. Respectively press 8, 2, 4, 6, the
cursor will move up, down, left and right.
5.2.10 Volt Step- can be set in BDV test mode
Volt Step ranges from 1% to 50%. The step voltage is calculated as following formula: