ー 61 ー
11-1. Specifications of the FLS-1000
11. SPECIFICATIONS
220V model
230V model
240V model
Equipment category: (medical/laboratory)
Laboratory equipment
Model name
Autoclave FLS-1000
Operating temperature
range
Sterilize
105℃ to 123℃
(It varies depending on altitude level.)
Maximum operating pressure
0.193MPa
Temperature
Display form
Digital
Display range
-15℃ to 175℃
Pressure gauge
Display form
Analog
Display range
0MPa to 0.4MPa (red zone:0.193MPa)
Heat source (electric heater)
3.0kW
3.0kW
3.3kW
Safety devices
●Safety valve
●No water heating prevention
●Leakage breaker
●Lid interlock
●Over-temperature prevention
●Over-pressure prevention
●Temperature sensor malfunction detection
●Pressure sensor malfunction detection
●Exhaust valve malfunction detection
●Low sterilizing water level detection
●Chamber lid open/close detection
●Drain access cover open detection
●Exhaust tank full level detection
Leakage breaker
Rated breaking current
15A
Rated sensitivity current
10mA
Protection against electric shock
Class I
Time
Display form
Digital
Display range
Sterilize
0:00
〜
9:59
/
0
〜
99hours/0
〜
999minutes/0:00
〜
9:59, 10
〜
99hours
(
switchable
)
Setting range
Sterilize
0:00
〜
9:59
/
0
〜
99hours/0
〜
999minutes/0:00
〜
9:59, 10
〜
99hours
(
switchable
)
Chamber dimensions
Inside diameter × inside dimension :
φ
450×678mm
Chamber capacity
100L(0.100m3
㎥
)
Effective diameter × effective depth
φ
440×501mm
Effective internal capacity
76L(0.076
㎥
)
Chamber material
SUS304 (stainless steel)
Main body dimensions(mm)
600W×600D×940H, top plate height: 785
680W×680D×1022H (including the pedal and control panel)
Net weight
108kg