ー 54 ー
1. Leakage breaker
Turn on the power supply switch on the main body.
Push the gray test button provided inside the power switch with a thin rod.
If the autoclave automatically switches off the power, the leakage breaker functions normally.
If it does not switch off, contact your dealer or our local representative.
<
To inspect each part
>
①
②
2. Chamber
Check chamber lid gasket for cracks or other damage.
Check chamber interior for corrosion, cracks or other damage.
Check chamber lid for corrosion, cracks or other damage.
Check hinges, hinge shafts, retaining rings, lid hooks and lid catches for corrosion, cracks or
other damage.
Check safety valve for any signs of corrosion, cracks or other damage. The safety valve can
be visually inspected through the inspection port by removing its cover with a screwdriver.
After inspection reinstall the cover.
①
②
③
④
⑤
3. Power plug
Unplug the autoclave.
Wipe off any dust or dirt on the surface of the power plug using a dry cloth.
※
If the power plug generates abnormal heat or has any burn marks on it, stop using the autoclave
and call for your dealer or our local representative.
※
If the power plug is too loose in the outlet, stop using the autoclave and call for your dealer or our
local representative.
①
②
4. Chamber lid handle
Check chamber lid handle for deformation or other damage.
①