When the sterilizing water change indicator is on
>
Drain water from the chamber and pour water into the chamber.
※
Refer to “5-11. Draining the Sterilizing Water from the Chamber”.
Press and hold down the CHECK key.
The buzzer will sound in approximately two seconds and the number of the operated cycles
after change will be reset.
※
To continue operation without pouring water into the chamber, press the CHECK key. Five
cycles can be performed.
①
②
6-10. Sterilizing Water Change Alert Function
When the set number of sterilization cycles is reached, the sterilizing water change indicator
lights up to notify the operator of the timing for water change to prevent the chamber from
getting rust by contaminated water inside. When the sterilizing water change indicator is
turned on, follow the instructions below for pouring water into the chamber. If the start key
is pressed without changing water while the sterilizing water change indicator is on, “A01”
appears in the time display and the autoclave enters into the standby mode. The number of
cycles which activates the water change alert function can be changed depending on the level of
contamination in water.
<
To set the change alert
>
STOP
MEMORY
STERILIZE
LIQUID
START
FUNCTION
CHECK
TIMER
EXT.
SENSOR
COOLING
FAN
EXHAUST SPEED
FAST
TIME
TEMPERATURE
TIMER
READY
STERILIZE
LID
LID
WATER
DRAIN
COVER
CHANGE WATER
°C
SLOW
STERILIZE
IN PROGRESS
Press the FUNCTION key while the IN PROGRESS
indicator is off. The time display flashes “F00”.
Press the time setting up and down arrow keys to
show “F12” in the time display. The temperature
display shows the number of cycles which activates
the alert function when it is reached.
Press the START key. The temperature display
starts to flash.
Press the temperature setting up and down arrow keys to show the desired number of cycles.
※
The factory default value is 40.
※
The setting range is between 5 and 80 cycles. When “− −” is selected, the alert function will not
activate.
Press and hold down the START key to finalize the setting.
The buzzer will sound in approximately 4 seconds and the display will return to normal
display.
①
②
③
④
⑤
STOP
MEMORY
STERILIZE
LIQUID
START
FUNCTION
CHECK
TIMER
EXT.
SENSOR
COOLING
FAN
EXHAUST SPEED
FAST
TIME
TEMPERATURE
TIMER
READY
STERILIZE
LID
LID
WATER
DRAIN
COVER
CHANGE WATER
°C
SLOW
STERILIZE
IN PROGRESS
Содержание FLS-1000
Страница 9: ...ー 8 ー MEMO ...
Страница 10: ...ー 9 ー 2 2 Parts Identifications FLS 1000 Main Unit ...
Страница 51: ...ー 50 ー MEMO ...