Error code
Possible cause
Remedy
E20
The pressure inside the chamber is high
when the power is turned on.
※
It exceeds 5.6kPa.
Wait until the chamber pressure drops to 0MPa
and then turn on the power.
E22
The chamber pressure rises because the
shape of the article to be sterilized serves to
prevent air from escaping.
※
The pressure exceeds the value predicted
by the chamber temperature by equal to
or greater than 25kPa during full opening
of the valve.
Turn off the power switch and wait until the
chamber pressure becomes 0MPa to open the
lid. And then take some measures to allow air to
escape from the load or reduce the volume of the
load.
(Refer to “5-6. Placing Articles to be Sterilized”.)
E23
The pressure drops because the exhaust
channel is blocked.
※
The pressure drops below the value
predicted by the chamber temperature by
equal to or less than 30kPa.
Turn off the power switch and wait until the
chamber pressure becomes 0MPa to open the
lid. Take out and place the load again so that the
hole in the chamber is not blocked. Open the
sterilization bag if it is closed.
E25
Control system malfunction
※
Memory abnormality
Unplug the power plug of the FLS-1000 from
the outlet and contact your dealer or our local
representative.
E26
Control system malfunction
※
Memory’s lifetime has expired.
Unplug the power plug of the FLS-1000 from
the outlet and contact your dealer or our local
representative.
E27
Control system malfunction
※
External temperature sensor for articles
to be sterilized is not adjusted.
Unplug the power plug of the FLS-1000 from
the outlet and contact your dealer or our local
representative.
E28
Control system malfunction
※
Chamber temperature sensor is not
adjusted.
Unplug the power plug of the FLS-1000 from
the outlet and contact your dealer or our local
representative.