ー 55 ー
9. TROUBLESHOOTING
If any system malfunction in the FLS-1000 is suspected, when it is operated properly according
to this Operator’s Manual, check the cause referring to the table below given in “9-1. Symptoms
and Solutions” and take the corresponding remedy. If the symptom of the malfunction is not
identified or the remedies described do not solve the problem, unplug the power plug of the
FLS-1000 from the outlet and contact your dealer or our local representative.
9-1. Symptoms and Solutions
Symptom
Possible cause
Remedy
No display appears on the control
panel when the power switch is
turned on.
FLS-1000 is not properly plugged.
Check the power cord for proper
connection.
Fuse or breaker at the power supply
side is blown or tripped.
Connect the power cord to the
power supply with the output
capacity required by the FLS-1000.
READY indicator does not turn
on.
Lid open indicator is on.
Check opening and closing of the lid
and then close the lid.
Low sterilizing water level indicator
is on.
Fill the chamber with water.
Drain indicator is on.
Drain the exhaust tank.
Drain access cover open indicator is
on.
Close the drain access cover.
Sterilizing water change indicator is
on.
Change water in the chamber.
Temperature rises slowly.
Extremely large amount of water is
in the chamber.
Check and adjust a level of water in
the chamber.
FLS-1000 is loaded with a large
amount of articles to be sterilized.
Check and adjust a volume of the
load.
Sterilization is not performed
sufficiently.
Sterilization time is insufficient.
Extend the sterilization time.
Steam leaks from the chamber lid. Chamber lid gasket or chamber
opening is covered with dust or dirt.
Clean the chamber lid gasket and
chamber opening.
Chamber lid gasket is damaged.
Unplug the power plug of the
FLS-1000 from the outlet and
contact your dealer or our local
representative.
Steam leaks from the safety valve. Chamber pressure is too high.
Unplug the power plug of the
FLS-1000 from the outlet and
contact your dealer or our local
representative.
The FLS-1000 won’t start or will
stop operation while displaying
an error code.
Take each remedy corresponding to the error code displayed, referring to
the “Error Code Table” given below.