SS-OCT System Base Unit
Chapter 6: Troubleshooting
Page 46
MTN013142-D02
6.1. Changing the Input Fuses
If for some reason you need to replace a broken fuse in the base unit, you must perform the following procedure:
Remove the AC input cable that may be connected to the unit.
Slide open the cover of the fuse holder located at the rear panel of the either the Laser Module or the
Imaging Module. Remove the existing fuse and install the appropriate replacement fuse for the
respective unit.
1. Use only IS 1A 250VAC Type T 5x20mm style fuses (IEC 60127-2/III, low breaking capacity,
slow blow) for the Base Unit.
Slide the fuse cover closed.
Figure 60
Fuse Cover on Rear Panel
If for some reason you need to replace the mains lead, the power connection is made using the IEC 60320 C14
plug. When choosing the cord, ensure to choose a cord following local applicable standards. The cord must be
specified for 10 A, 250 V.
To avoid electrical shock, the power cord protective ground conductor must be connected to ground.