![THORLABS VEG200 Series Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/thorlabs/veg200-series/veg200-series_user-manual_1111028054.webp)
SS-OCT System Base Unit
Chapter 9: Specifications
Page 50
MTN013142-D02
Chapter 9
Specifications
General Performance Specifications – Vega Series Base Unit
Supply Voltage for Base Unit
*
100 V – 240 V / AC
Maximum Power Consumption
100 W
Weight of Base Unit (Approx.)
11.5 kg
Storage/Operating Temperature
10 °C to 35 °C
Dimensions of OCT-Stand (L x W x H)
206 mm x 305 mm x 248 mm
Dimensions of Base Unit (L x W x H)
361.8 mm x 305.0 mm x 143.0 mm
Airborne Noise Emission
<70 dB
A
*
Base unit has universal AC input.
Table 4 General Specifications
Optical Performance Specifications – Vega Series Base Unit
Base Unit
VEG210
VEG220
Central Wavelength
1300 nm
1300 nm
Axial Scan Rate
100 kHz
200 kHz
Minimum Pixels per A-Scan
1312
960
Sensitivity
102 dB (with OCT-LK4)
98 dB (with OCT-LK4)
Maximum Imaging Depth
Air/Water (typical)
11 mm / 8.3 mm
8.0 mm / 6.0 mm
Axial Resolution Air/Water (typical)
16 µm / 12 µm
16 µm / 12 µm
Lateral Resolution at Focus with OCT Scan Lens Kit
20 µm (OCT-LK4)
20 µm (OCT-LK4)
Table 5 Vega Specifications