PM320E

7.2   Exchange Mains Fuses

The two power input fuses are externally accessible. If they have opened due to line distortions,
incorrect line voltage or other causes, they can be replaced from the rear without opening the
unit. 

 Attention 

To avoid risk of fire the appropriate fuses for the corresponding line voltage must be used. 

1. Turn off the PM320E and disconnect the mains cable. 

2. The  fuse  holder  (see  figure  in  section  

Line  Voltage  Setting

 is  located  below  the  3-pole

power connector of the mains jack. Release the fuse holder by pressing its plastic retainers
with the aid of a small screwdriver. The retainers are located on the right and left side of the
holder and must be pressed towards the center.

3. Replace the defective or wrong fuses and press in the fuse holder until locked on both sides.

Take care to maintain the correct rotation of the line voltage indicator / switch which contains
the left fuse and is plugged into the fuse holder. The appropriate line voltage marking must
be visible in the window of the fuse holder.

Fuse types

100 V

1000 mA, time-lag, 250V

T1A250V

115 V

1000 mA, time-lag, 250V

T1A250V

230 V

500 mA, time-lag, 250V

T0.5A250V

All  fuses  must  meet  IEC  specification  60127-2/III,  time  characteristic:  time-lag  (T),  250V  AC,
size 5 x 20 mm.

Содержание PM320E

Страница 1: ...Thorlabs Instrumentation Optical Power and Energy Meter PM320E Operation Manual 2016...

Страница 2: ...Copyright 2016 Thorlabs GmbH Version Date Item No 2 1 10 Mar 2016 M0009 510 618...

Страница 3: ...easurement 15 3 1 6 Setting an Attenuation Factor 15 3 1 7 Trigger Input 16 3 2 Photodiode Power Sensor Operation 16 3 2 1 Wavelength Correction 17 3 2 2 Dark Current Adjustment 17 3 2 3 Bandwidth Set...

Страница 4: ...52 5 9 Error Reporting 6 Measurement Considerations 54 54 6 1 Choosing the Right Sensor 54 6 2 Reducing Noise for High Accuracy Measurements 54 6 3 Power Measurement of Pulsed Signals 55 6 4 Line Wid...

Страница 5: ...67 8 6 Thorlabs End of Life Policy 68 8 7 Thorlabs Worldwide Contacts...

Страница 6: ...international partners are looking forward to hearing from you Thorlabs GmbH Warning Sections marked by this symbol explain dangers that might result in personal injury or death Always read the associ...

Страница 7: ...age setting see section Line Voltage Setting and the mains fuses see section Exchange Mains Fuses The PM320E must not be operated in explosion endangered environments Do not remove covers Do not obstr...

Страница 8: ...re not to be used within the range of three meters of this unit since the electromagnetic field intensity may then exceed the maximum allowed disturbance values according to IEC 61326 1 This product h...

Страница 9: ...aration Attention Prior to switching on your PM320E please check if the line voltage setting see voltage selector at the rear panel corresponds to your mains voltage and make sure the correct fuses ar...

Страница 10: ...labs GmbH 8 PM320E 2 3 Operating Elements 2 3 1 Front Panel 1 Display 2 Control Knob 3 Power Switch 4 Soft Keys 5 6 7 Display Selector CH1 8 Photo Diode Input CH1 9 Photo Diode Input CH2 10 Display Se...

Страница 11: ...nd Connector Banana Jack 4 mm 3 USB Interface Connector Type B 4 MISC Output Connector BNC 5 CH1 Sensor Connector DSUB 9 Pin female 6 Analog Output CH1 BNC 7 CH2 Sensor Connector DSUB 9 Pin female 8 A...

Страница 12: ...2016 Thorlabs GmbH 10 PM320E 2 3 3 Main Displays Main Display Single Channel CH1 Main Display Dual Channel Display Power Trend Graph CH2 Display Power Statistics CH2...

Страница 13: ...2016 Thorlabs GmbH 2 Getting Started 11 Display Energy Statistics CH2 Display Energy Histogram CH1...

Страница 14: ...nd Statistics ON ON Channel 1 Channel 2 Dual Channel Screen To access the statistics and graphics displays select a single channel screen and toggle the right soft button CHx Display 3 1 2 System Sett...

Страница 15: ...the information of the current display Sensor Main Display CH 1 CH2 Sub Display CH1 CH 2 Photodiode Power W Power dBm Delta Power W Delta Power dB Ratio CH1 CH2 CH2 CH1 dB Power Difference CH1 CH2 CH...

Страница 16: ...OK soft key When power meter sensors are connected the ranges will show in Watts with energy sensors in Joules For the photodiode inputs the ranges can additionally be set in Amperes for the thermal...

Страница 17: ...setup menu 3 1 7 Trigger Input For power and energy measurements the PM320E has two trigger modes that can be selected for each channel in the CHx Setup under Trigger Source Depending on the connecte...

Страница 18: ...ensity filters and also integrating sphere materials show a wavelength dependent behavior and therefore each sensor is individually calibrated over the spectral working range in 5 nm steps The individ...

Страница 19: ...light To perform the adjustment cover the sensor with the metal protection cap or when a fiber connector is assembled plug a fiber laser switched off and press the Control knob or press the Enter sof...

Страница 20: ...the active area thermal disk and the sensor heat sink with no light incident to the active area or when the heat sink gets hot under light exposure An indicator that the sensor needs to be zeroed is...

Страница 21: ...ent to one of the DB 9 connectors CH1 or CH2 with pin 8 to plus pole and pin 3 to ground Connect a resistor 1 kW to 10 kW between pin 7 PRESENT and 3 GND The PM320E will recognize a Thermal Head Adapt...

Страница 22: ...play will be held on the last measured value To reset the reading scroll to the Reset setting and press the multi control knob The measurement value will show until the next pulse appears Pyroelectric...

Страница 23: ...power or energy density in the Main Display For entering the beam diameter navigate the CHx Setup to Beam Diameter 1 e2 3 4 4 Custom Pyroelectric Sensors Beside Thorlabs pyroelectric energy sensors a...

Страница 24: ...polarity can be changed only with the Bias voltage switched off 3 5 2 Setting the Photodiode Responsivity The photo current can be displayed as a power level in Watts or dBm only if the polarity has b...

Страница 25: ...ut in the rear panel To start the data acquisition press the Start Reset button Statistics Mode for Power Energy Measurements 3 7 Chart Mode Power Sensors The chart mode can be accessed by toggling th...

Страница 26: ...left and right side of the histogram 3 9 System Configuration The system configuration menu contains the following items Item Value Description LCD Brightness 0 100 Sets Backlight of the LCD Sound ON...

Страница 27: ...t that provides a voltage from 10V to 10V depending on the selected function The voltage from this output can be configured in the Misc Output Configuration in the main Menu in terms of each possible...

Страница 28: ...the installation you also need to install NI VISA form the distribution CD or from the National Instruments web site After the software is installed successfully connect the PM320E to a USB port of yo...

Страница 29: ...ins the following items Thorlabs Vendor ID 0x1313 Product ID 0x8022 PM320E Serial number The utility software will recognize and update connected sensors while the program is running The display data...

Страница 30: ...s Enable query IEEE488 2 1992 10 11 ESR Standard Event Status Register query IEEE488 2 1992 10 12 CLS Clear Status command IEEE488 2 1992 10 3 Descriptions Identification Query Command format IDN Resp...

Страница 31: ...s Status Byte see also IEEE488 2 1992 10 36 and chapter Status Reporting in this document Standard Event Status Enable Command Command format ESE DECIMAL NUMERIC PROGRAM DATA Description Sets the devi...

Страница 32: ...l 1 2 measured photodiode current value A FETCH VOLT 1 2 VAL Fetch the channel 1 2 measured photodiode thermopile voltage value V FETCH UPY 1 2 VAL Fetch the channel 1 2 measured pyroelectric adapter...

Страница 33: ...NUMERIC RESPONSE DATA Description Queries the channel 1 2 measured pyroelectric adapter voltage in Volt V Note If an overflow occurs the value Inf or Inf is returned Query the Zero Offset Value Comma...

Страница 34: ...he selected channel The status of a measurement can be checked with the MEAS CHECK 1 2 command see below Abort a running measurement Command syntax MEAS ABORT 1 2 Description Cancels a running measure...

Страница 35: ...ency NR3 NUMERIC RESPONSE DATA Description Queries the channel 1 2 measured pulse repitition frequency value in Hertz Hz of the measurement started by MEAS INIT 1 2 Note If an overflow occurs the valu...

Страница 36: ...lectric voltage range VRANGE 1 2 MAX Query the channel 1 2 maximum settable pyroelectric voltage range WAVEL 1 2 VAL Set the wavelength for channel 1 2 WAVEL 1 2 VAL Query the currently used wavelengt...

Страница 37: ...nsor trigger level range TRIGMODE 1 2 Set the channel 1 2 trigger mode TRIGMODE 1 2 Query the channel 1 2 trigger mode BEAMDIA 1 2 VAL Set the pyro sensor beam diameter 1 2 BEAMDIA 1 2 VAL Query the p...

Страница 38: ...uery Channel 1 2 Maximum Current Range Command format IRANGE 1 2 MAX Response format CHARACTER RESPONSE DATA AUTO R100NA R1UA R10UA R100UA R1MA R10MA Description Queries the channel 1 2 maximum settab...

Страница 39: ...ponse format VRANGE 1 2 CHARACTER PROGRAM DATA R100MV R1V R10V R100V Description Sets the pyroelectric voltage range for channel 1 2 Query Channel 1 2 Pyroelectric Voltage Range Command format VRANGE...

Страница 40: ...Description Queries channel 1 2 power attenuation range Set Channel 1 2 Photodiode Input Sensor Bandwidth Command format PDBANDW 1 2 CHARACTER PROGRAM DATA LOW HIGH Description Sets the photo diode i...

Страница 41: ...MERIC RESPONSE DATA Description Queries the channel 1 2 currently used photodiode input BIAS voltage V Query Channel 1 2 Photodiode Input BIAS Voltage Range Command format PDBIAS 1 2 RNG Response form...

Страница 42: ...esponsivity V W NR3 NUMERIC RESPONSE DATA Description Queries channel 1 2 thermopile responsivity range Set Channel 1 2 Pyro Sensor Responsivity Command format PYRESP 1 2 VAL DECIMAL NUMERIC PROGRAM D...

Страница 43: ...ommand format BEAMDIA 1 2 RNG Response format minimum settable pyro sensor beam diam m NR3 NUMERIC RESPONSE DATA maximum settable pyro sensor beam diam m NR3 NUMERIC RESPONSE DATA currently used pyro...

Страница 44: ...nitor 30201 Thermopile Adaptor 30202 Thermopile Sensor 30212 Thermopile Sensor with temperature monitor 30301 Pyroelectric Adaptor 30302 Pyroelectric Sensor 30312 Pyroelectric Sensor with temperature...

Страница 45: ...OUT OFFS VAL Query the MISC Output offset voltage MISCOUT OFFS RNG Query the MISC Output offset voltage range Descriptions Set MISC Output Operating Mode Command format MISCOUT MODE CHARACTER PROGRAM...

Страница 46: ...1V 3mW R10MW 1V 10mW R30MW 1V 30mW R100MW 1V 100mW R300MW 1V 300mW R1W 1V 1W R3W 1V 3W R10W 1V 10W R30W 1V 30W R100W 1V 100W Query MISC Output sensitivity Watt modes Command format MISCOUT SENS WATT...

Страница 47: ...R10DB 1V 10dB Query MISC Output sensitivity dB modes Command format MISCOUT SENS DB Response format CHARACTER RESPONSE DATA R0_1DB R0_3DB R1DB R3DB R10DB Description Queries the MISC Output sensitivit...

Страница 48: ...MAL NUMERIC PROGRAM DATA Description Sets the MISC Output offset voltage V Query MISC Output Offset Voltage Command format MISCOUT OFFS VAL Response format offset voltage NR3 NUMERIC RESPONSE DATA Des...

Страница 49: ...STAT ERR EVT Response format NR1 NUMERIC RESPONSE DATA Description Queries the device s Error Event Register see also chapter Status Reporting in this document Set Device Error Event Enable Register...

Страница 50: ...Sets the device s Operation Event Enable Register see also chapter chapter Status Reporting in this document Query Device Operation Event Enable Register Command format STAT OPER ENA Response format...

Страница 51: ...s the device s hardware option bitmap software option bitmap and the mainboard ID For further information see chapters Hardware Options Software Options and Supported Mainboard IDs in this document Qu...

Страница 52: ...ce information Set Linefilter Command format SYST LINEFILTER CHARACTER PROGRAM DATA 50HZ 60HZ Description Sets the device line filter setting Note For optimum device operation set the line filter to y...

Страница 53: ...ne error in the error queue 1 DOS Device Operation Status Structure Summary Bit 0 Reserved Device Error Status Structure The Device Error Status Structure reflects asynchronous device errors Bits in t...

Страница 54: ...eading keyboard controller signature 16 Keyboard controller update failed with timeout at writing image 17 Keyboard controller update failed with timeout at reading back image 18 Keyboard controller u...

Страница 55: ...running 82 Zero procedure failed 83 No PD polarity switching during BIAS on 100 Authentication required for operation 101 Authentication failed 110 Operation is not allowed in SERVICE MODE 111 Operat...

Страница 56: ...be linked to earth ground e g via a post energy sensors should be mounted isolated because the housing is connected to the PM320E analog ground The sensor cable must handle very small current or volta...

Страница 57: ...show the right optical power for a symmetrical spectral response shape 6 5 Temperature Effects on Thermal Sensors Thermal sensors react on any temperature differences that occur between thermal disc a...

Страница 58: ...acceptance angle is calculated by the numerical aperture of the fiber and is defined as the inverse sine of the numerical aperture Fore typical single mode fibers the total angle of the emitted light...

Страница 59: ...10 10 90 V 115 V or 115 V 10 10 104 V 132 V or 230 V 10 10 207 V 264 V The line voltage setting can be changed on the rear without opening the unit 1 Turn off the PM320E and disconnect the mains cabl...

Страница 60: ...fuse holder by pressing its plastic retainers with the aid of a small screwdriver The retainers are located on the right and left side of the holder and must be pressed towards the center 3 Replace t...

Страница 61: ...nect the PM320E to an USB port of your PC and launch the DFU wizard from the start bar Follow the wizard instructions Please refer to www thorlabs com for the latest PM320E firmware version that can b...

Страница 62: ...mal sensor is connected Check if the channel display is in the right mode to display absolute power Conduct the dark current adjustment procedure A photodiode is connected Check the polarity setting I...

Страница 63: ...A 10 pA 1 x 107 V A 0 5 f s 1 kHz 25 Hz Voltage Input Thermal Sensors Full Scale Measurement Ranges 1 mV 1 V in decade steps Maximum Display Resolution 0 1 V Display Units V W dBm Input Impedance 1 MW...

Страница 64: ...S3xxC Thermal Sensors ES1xxC ES2xxC Pyroelectric Sensors Anode and Cathode Grounded Photodiodes Custom Thermal Sensors Custom Pyroelectric Sensors General Data Line Voltage selectable 100 V 115 V 230...

Страница 65: ...on This pin is uniquely used for the memory in the sensor head and must not be used Connecting this pin may cause malfunction of the PM320E 3 photodiode ground anode thermal and pyro sensor ground ana...

Страница 66: ...61000 3 2 AC Power Line Harmonic Emissions IEC 61000 3 3 Voltage Fluctuations and Flicker FCC EMC Compliance Emissions comply with the Class B Limits of FCC Code of Federal Regulations 47 Part 15 Subp...

Страница 67: ...and software determined by Thorlabs GmbH for this unit to operate fault free provided that they are handled according to our requirements However Thorlabs GmbH does not warrant a fault free and uninte...

Страница 68: ...oduct In no event shall any liability exceed the purchase price of the product Please note that the content of this User Manual is neither part of any previous or existing agreement promise representa...

Страница 69: ...ce does not refer to other Thorlabs GmbH products such as pure OEM products that means assemblies to be built into a unit by the user e g OEM laser driver cards components mechanics and optics left ov...

Страница 70: ...horlabs Sweden AB M lndalsv gen 3 412 63 G teborg Sweden Tel 46 31 733 30 00 Fax 46 31 703 40 45 www thorlabs com Email scandinavia thorlabs com France Thorlabs SAS 109 rue des C tes 78600 Maisons Laf...

Отзывы: