© 2016 Thorlabs GmbH
5 Command Reference
35
Command
Description
:THTIMECONST[1,2]:VAL
Set the channel 1/2 thermopile adapter accelerator time constant value [s]
:THTIMECONST[1,2]:VAL?
Query the channel 1/2 thermopile adapter accelerator time constant value
[s]
:THTIMECONST[1,2]:RNG?
Query the channel 1/2 thermopile adapter accelerator time constant value
range [s]
:THRESP[1,2]:VAL
Set the thermopile responsivity for channel 1/2
:THRESP[1,2]:VAL?
Query the currently used thermopile responsivity for channel 1/2
:THRESP[1,2]:RNG?
Query the channel 1/2 thermopile responsivity range
:PYRESP[1,2]:VAL
Set the pyro sensor responsivity for channel 1/2
:PYRESP[1,2]:VAL?
Query the pyro sensor responsivity for channel 1/2
:PYRESP[1,2]:RNG?
Query the channel 1/2 pyro sensor responsivity range
:TRIGLVL[1,2]:VAL
Set the pyro sensor trigger level for channel 1/2
:TRIGLVL[1,2]:VAL?
Query the pyro sensor trigger level for channel 1/2
:TRIGLVL[1,2]:RNG?
Query the channel 1/2 pyro sensor trigger level range
:TRIGMODE[1,2]
Set the channel 1/2 trigger mode
:TRIGMODE[1,2]?
Query the channel 1/2 trigger mode
:BEAMDIA[1,2]:VAL
Set the pyro sensor beam diameter 1/2
:BEAMDIA[1,2]:VAL?
Query the pyro sensor beam diameter for channel 1/2
:BEAMDIA[1,2]:RNG?
Query the channel 1/2 pyro sensor beam diameter range
:ADAPTYPE[1,2]
Set the sensor type to use if an adapter is connected
:ADAPTYPE[1,2]?
Query the sensor type to use if an adapter is connected
Descriptions
Set Channel 1/2 Power Range
Command format:
:PRANGE[1,2] <CHARACTER PROGRAM DATA>
[AUTO,R1NW,R10NW,R100NW,R1UW,R10UW,R100UW,R1MW,R10MW,
R100MW,R1W,R10W,R100W,R1KW]
Description:
Sets the power range for channel 1/2.
Query Channel 1/2 Power Range
Command format:
:PRANGE[1,2]?
Response format:
<CHARACTER RESPONSE DATA>
[AUTO,R1NW,R10NW,R100NW,R1UW,R10UW,R100UW,R1MW,R10MW,
R100MW,R1W,R10W,R100W,R1KW]
Description:
Queries the channel 1/2 power range.
Query Channel 1/2 Minimum Power Range
Command format:
:PRANGE[1,2]:MIN?
Response format:
<CHARACTER RESPONSE DATA>
[AUTO,R1NW,R10NW,R100NW,R1UW,R10UW,R100UW,R1MW,R10MW,
R100MW,R1W,R10W,R100W,R1KW]
Description:
Queries the channel 1/2 minimum settable power range.
Содержание PM320E
Страница 1: ...Thorlabs Instrumentation Optical Power and Energy Meter PM320E Operation Manual 2016...
Страница 2: ...Copyright 2016 Thorlabs GmbH Version Date Item No 2 1 10 Mar 2016 M0009 510 618...
Страница 5: ...67 8 6 Thorlabs End of Life Policy 68 8 7 Thorlabs Worldwide Contacts...
Страница 13: ...2016 Thorlabs GmbH 2 Getting Started 11 Display Energy Statistics CH2 Display Energy Histogram CH1...