5.2 Measurement Commands
Command List
Command
Description
:AVER[1,2]?
Query the channel 1/2 averaging rate
:AVER[1,2]
Set the channel 1/2 averaging rate
:POW[1,2]:VAL?
Query the channel 1/2 measured power [W]
:CURR[1,2]:VAL?
Query the channel 1/2 measured current [A]
:VOLT[1,2]:VAL?
Query the channel 1/2 measured voltage [V]
:UPY[1,2]:VAL?
Query the channel 1/2 measured pyroelectric adapter voltage [V]
:ZERO[1,2]:VAL?
Query the channel 1/2 Zero Offset Value
:ZERO[1,2]:ADJ
Performs a Zero Offset Value adjustment for channel 1/2
:ENERGY[1,2]:VAL?
Query the channel 1/2 measured pulse energy [J]
:PAVG[1,2]:VAL?
Query the channel 1/2 measured average power [W]
:FREP[1,2]:VAL?
Query the channel 1/2 measured pulse rep. frequency [Hz]
:EDENSITY[1,2]:VAL?
Query the channel 1/2 measured energy density [J/cm²]
:PDENSITY[1,2]:VAL?
Query the channel 1/2 measured power density [W/cm²]
:MEAS:INIT[1,2]
Initiate a measurement on channel 1/2
:MEAS:ABORT[1,2]
Abort a measurement initiated on channel 1/2
:MEAS:CHECK[1,2]?
Check if measurement initiated on channel 1/2 has finished
:FETCH:CURR[1,2]:VAL?
Fetch the channel 1/2 measured photodiode current value [A]
:FETCH:VOLT[1,2]:VAL?
Fetch the channel 1/2 measured photodiode thermopile voltage value [V]
:FETCH:UPY[1,2]:VAL?
Fetch the channel 1/2 measured pyroelectric adapter voltage value [V]
:FETCH:POW[1,2]:VAL?
Fetch the channel 1/2 measured optical power [W]
:FETCH:ENERGY[1,2]:VAL?
Fetch the channel 1/2 measured pulse energy value [J]
:FETCH:FREP[1,2]:VAL?
Fetch the channel 1/2 measured pulse repetition frequency value [Hz]
:FETCH:EDENSITY[1,2]:VAL?
Fetch the channel 1/2 measured pulse energy density value [J/cm²]
:FETCH:PAVG[1,2]:VAL?
Fetch the channel 1/2 measured average power value [W]
:FETCH:PDENSITY[1,2]:VAL?
Fetch the channel 1/2 measured power density value [W/cm²]
Descriptions
Query the Averaging Rate
Command format:
:AVER[1,2]?
Response format:
currently used Averaging Rate
<NR1 NUMERIC RESPONSE DATA>
Description:
Queries the currently used averaging rate for channel 1/2.
Set the Averaging Rate
Command format:
:AVER[1,2] <DECIMAL NUMERIC PROGRAM DATA>
Description:
Sets the averaging rate for channel 1/2.
Note: The averaging rate does not influence the front panel display. The instrument
measures at 500Hz rate. Use this value to achieve an appropriate line filtering. To
achieve a good 60Hz filtering use a multiple of 25. To achieve a good 50Hz filtering use
a multiple of 10.
Содержание PM320E
Страница 1: ...Thorlabs Instrumentation Optical Power and Energy Meter PM320E Operation Manual 2016...
Страница 2: ...Copyright 2016 Thorlabs GmbH Version Date Item No 2 1 10 Mar 2016 M0009 510 618...
Страница 5: ...67 8 6 Thorlabs End of Life Policy 68 8 7 Thorlabs Worldwide Contacts...
Страница 13: ...2016 Thorlabs GmbH 2 Getting Started 11 Display Energy Statistics CH2 Display Energy Histogram CH1...