3 Operating Instruction
15
3.1.6 Setting an Attenuation Factor
The Attenuation Menu allows to set a user correction factor to the power reading. This is very
convenient, for example, to compensate the measurement result for the attenuation of a filter or
beam splitter in the system.
To enter an attenuation factor, press the <CHx Setup> soft button, navigate and click the
‘Attenuation’ button. Set the desired attenuation in dB, then press <OK> or click the control
button. Exit the setup menu.
3.1.7 Trigger Input
For power and energy measurements the PM320E has two trigger modes that can be selected
for each channel in the CHx Setup under <Trigger Source>:
Depending on the connected sensor the selected trigger source has the function shown in the
table below.
Sensor
Trigger Source Auto
Trigger Source Extern
Photodiode /
Thermal
continuous power measurement
with fixed sampling rate
A rising edge TTL signal at the rear
BNC trigger input will trigger a single
measurement
Pyroelectric
All pulses that exceed the trigger
threshold will be measured
automatically
A rising edge TTL signal resets the
peak detector, the system is waiting
for the next incoming pulse to
measure
Содержание PM320E
Страница 1: ...Thorlabs Instrumentation Optical Power and Energy Meter PM320E Operation Manual 2016...
Страница 2: ...Copyright 2016 Thorlabs GmbH Version Date Item No 2 1 10 Mar 2016 M0009 510 618...
Страница 5: ...67 8 6 Thorlabs End of Life Policy 68 8 7 Thorlabs Worldwide Contacts...
Страница 13: ...2016 Thorlabs GmbH 2 Getting Started 11 Display Energy Statistics CH2 Display Energy Histogram CH1...