![THORLABS PM320E Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/thorlabs/pm320e/pm320e_operation-manual_1110983040.webp)
© 2016 Thorlabs GmbH
38
PM320E
Query Channel 1/2 Wavelength Range
Command format:
:WAVEL[1,2]:RNG?
Response format:
minimum settable wavelength [nm]
<NR3 NUMERIC RESPONSE DATA>,
maximum settable wavelength [nm]
<NR3 NUMERIC RESPONSE DATA>,
currently used wavelength [nm]
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries channel 1/2 wavelength range.
Set Channel 1/2 Power Attenuation
Command format:
:ATTEN[1,2]:VAL <DECIMAL NUMERIC PROGRAM DATA>
Description:
Sets the channel 1/2 power attenuation [dB].
Query Channel 1/2 Power Attenuation
Command format:
:ATTEN[1,2]:VAL?
Response format:
currently used power attenuation
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries the channel 1/2 currently used power attenuation [dB].
Query Channel 1/2 Power Attenuation Range
Command format:
:ATTEN[1,2]:RNG?
Response format:
minimum settable power attenuation [dB]
<NR3 NUMERIC RESPONSE DATA>,
maximum settable power attenuation [dB]
<NR3 NUMERIC RESPONSE DATA>,
default power attenuation [dB]
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries channel 1/2 power attenuation range.
Set Channel 1/2 Photodiode Input/Sensor Bandwidth
Command format:
:PDBANDW[1,2] <CHARACTER PROGRAM DATA> [LOW,HIGH]
Description:
Sets the photo diode input/sensor bandwidth setting.
Query Channel 1/2 Photodiode Input/Sensor Bandwidth
Command format:
:PDBANDW[1,2]?
Response format:
<CHARACTER RESPONSE DATA> [LOW,HIGH]
Description:
Queries the photo diode input/sensor bandwidth setting.
Set Channel 1/2 Photodiode Input Polarity
Command format:
:PDPOL[1,2] <CHARACTER PROGRAM DATA> [AG,CG]
Description:
Sets the photodiode input polatity to anode ground/cathode ground..
Query Channel 1/2 Photodiode Input Polarity
Command format:
:PDPOL[1,2]?
Response format:
<CHARACTER RESPONSE DATA> [AG,CG]
Description:
Queries the photodiode input polatity.
Set Channel 1/2 Photodiode Responsivity
Command format:
:PDRESP[1,2]:VAL <DECIMAL NUMERIC PROGRAM DATA>
Description:
Sets the channel 1/2 photodiode responsivity [A/W].
Query Channel 1/2 Photodiode Responsivity
Command format:
:PDRESP[1,2]:VAL?
Response format:
currently used photodiode responsivity
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries the channel 1/2 currently used photodiode responsivity [A/W].
Содержание PM320E
Страница 1: ...Thorlabs Instrumentation Optical Power and Energy Meter PM320E Operation Manual 2016...
Страница 2: ...Copyright 2016 Thorlabs GmbH Version Date Item No 2 1 10 Mar 2016 M0009 510 618...
Страница 5: ...67 8 6 Thorlabs End of Life Policy 68 8 7 Thorlabs Worldwide Contacts...
Страница 13: ...2016 Thorlabs GmbH 2 Getting Started 11 Display Energy Statistics CH2 Display Energy Histogram CH1...