5 Command Reference
39
Query Channel 1/2 Photodiode Responsivity Range
Command format:
:PDRESP[1,2]:RNG?
Response format:
minimum settable photodiode resp. [A/W]
<NR3 NUMERIC RESPONSE DATA>,
maximum settable photodiode resp. [A/W]
<NR3 NUMERIC RESPONSE DATA>,
currently
used photodiode responsivity [A/W]
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries channel 1/2 photodiode responsivity range.
Set Channel 1/2 Photodiode Input BIAS State
Command format:
:PDBIAS[1,2] <CHARACTER PROGRAM DATA> [ON,OFF]
Description:
Switches the photodiode input BIAS voltage on/off..
Query Channel 1/2 Photodiode Input BIAS State
Command format:
:PDBIAS[1,2]?
Response format:
<CHARACTER RESPONSE DATA> [ON,OFF]
Description:
Queries the photodiode input BIAS state.
Set Channel 1/2 Photodiode Input BIAS Voltage
Command format:
:PDBIAS[1,2]:VAL <DECIMAL NUMERIC PROGRAM DATA>
Description:
Sets the channel 1/2 photodiode input BIAS voltage [V].
Query Channel 1/2 Photodiode Input BIAS Voltage
Command format:
:PDBIAS[1,2]:VAL?
Response format:
currently used BIAS voltage
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries the channel 1/2 currently used photodiode input BIAS voltage [V].
Query Channel 1/2 Photodiode Input BIAS Voltage Range
Command format:
:PDBIAS[1,2]:RNG?
Response format:
minimum settable BIAS voltage [V]
<NR3 NUMERIC RESPONSE DATA>,
maximum settable BIAS voltage [V]
<NR3 NUMERIC RESPONSE DATA>,
default BIAS voltage [V]
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries channel 1/2 photodiode input BIAS voltage range.
Set Channel 1/2 Thermopile Sensor Accelerator State
Command format:
:THACCEL[1,2] <CHARACTER PROGRAM DATA> [OFF,ON]
Description:
Sets the thermopile sensor accelerator state.
Query Channel 1/2 Thermopile Sensor Accelerator State
Command format:
:THACCEL[1,2]?
Response format:
<CHARACTER RESPONSE DATA> [OFF,ON]
Description:
Queries the channel 1/2 thermopile sensor accelerator state.
Set Channel 1/2 Thermal Time Constant
Command format:
:THTIMECONST[1,2]:VAL <DECIMAL NUMERIC PROGRAM DATA>
Description:
Sets the channel 1/2 thermal time constant [s]. (used only with thermopile adaptors)
Query Channel 1/2 Thermal Time Constant
Command format:
:THTIMECONST[1,2]:VAL?
Response format:
currently used thermal time constant
<NR3 NUMERIC RESPONSE DATA>
Description:
Queries the channel 1/2 currently used thermal time constant [s].
Содержание PM320E
Страница 1: ...Thorlabs Instrumentation Optical Power and Energy Meter PM320E Operation Manual 2016...
Страница 2: ...Copyright 2016 Thorlabs GmbH Version Date Item No 2 1 10 Mar 2016 M0009 510 618...
Страница 5: ...67 8 6 Thorlabs End of Life Policy 68 8 7 Thorlabs Worldwide Contacts...
Страница 13: ...2016 Thorlabs GmbH 2 Getting Started 11 Display Energy Statistics CH2 Display Energy Histogram CH1...