5 Command Reference
51
5.8 Status Reporting
Status Byte Register
The Status Byte Register gives a summary of all underlying status structures. The structure is
defined in IEEE488.2-1992-§11.2.
Bit #
Mnemonic
Description
7
reserved
6
RQS/MSS
Request Service / Master Summary Status
5
ESB
Standard Event Status Bit
4
MAV
Message Available. There is response data available for readout
3
DES
Device Error Status Structure Summary Bit
2
EAV
Error Available. There is at least one error in the error queue.
1
DOS
Device Operation Status Structure Summary Bit
0
Reserved
Device Error Status Structure
The Device Error Status Structure reflects asynchronous device errors. Bits in the according
event register are rising and falling edge triggered.
Bit #
Mnemonic
Description
15..9
reserved
8
SM2
Sensor missing channel 2
7..1
reserved
0
SM1
Sensor missing channel 1
Device Operation Status Structure
The Device Operation Status Structure reflects device states. Bits in the according event
register are rising edge and falling edge triggered.
Bit #
Mnemonic
Description
15..9
reserved
8
ZR2
Zero procedure running channel 2
7..1
reserved
0
ZR1
Zero procedure running channel 1
Содержание PM320E
Страница 1: ...Thorlabs Instrumentation Optical Power and Energy Meter PM320E Operation Manual 2016...
Страница 2: ...Copyright 2016 Thorlabs GmbH Version Date Item No 2 1 10 Mar 2016 M0009 510 618...
Страница 5: ...67 8 6 Thorlabs End of Life Policy 68 8 7 Thorlabs Worldwide Contacts...
Страница 13: ...2016 Thorlabs GmbH 2 Getting Started 11 Display Energy Statistics CH2 Display Energy Histogram CH1...