![THORLABS LPS710/M Скачать руководство пользователя страница 39](http://html.mh-extra.com/html/thorlabs/lps710-m/lps710-m_user-manual_1110767039.webp)
39
LPS710E Long Travel Piezo Stage and Controller
Chapter 5 Software Reference
5.1 GUI Panel
The following screen shot shows the graphical user interface (GUI) (one panel per
card fitted) displayed when accessing the controller using the APTUser utility.
Fig. 5.1 PPC001 Piezo Driver Software GUI
Output Control
- used to adjust and set the voltage or position output as displayed in
the main digital display. If the channel is in closed loop mode (set using the 'closed
loop' button on the panel) then this control can be used to adjust position output. If
open loop mode is selected (using the 'open loop' button) then the control is used to
adjust the voltage output. Note that the read out in the main digital display is
independent of the operating mode (open or closed loop) selected for the particular
channel. If the 'Volts' button is selected, then the read out is in volts, even if the channel
is in closed loop mode and the Output control is being used to adjust position.
Similarly, if the 'position' button is selected, the read out is in microns, derived from the
feedback signal, even if the channel is operating in open loop.
Jog
- used to increment or decrement the piezo position. When the button is clicked,
the piezo is driven in the selected direction, one step per click. The step size can be
set in the 'Settings' panel or by using the SetJogStepSize method.
Note
The serial number of the PPC001 unit associated with the GUI panel, the
APT server version number, and the version number (in brackets) of the
embedded software running on the unit, are displayed in the top right
hand corner. This information should always be provided when
requesting customer support.