58
Commands
_________________________________________________________________________________
______________________________________________________________________________________________________
© 2010 Thorlabs
3.2.27 Reading the laser diode voltage (VLD)
Programming:
":VLD:MEAS <NR1>"
Programs V
LD
e a measurement value for “ELCH” on
position <NR1> (1....8) in the output string.
Reading:
":VLD:ACT?"
Reads the actual laser voltage
[:VLD:ACT
<NR3><LF>]
":VLD:MIN_R?"
Reads U
LD
ADC = 0000
[:VLD:MIN_R
<NR3><LF>]
":VLD:MAX_R?"
Reads U
LD
– ADC = FFFF
[:VLD:MAX_R
<NR3><LF>]
":VLD:MEAS?"
Reads the position of V
LD
the “ELCH” output string
(
1....8, 0
if not selected)
[:VLD:MEAS
<NR1><LF>]
3.2.28 Reading the TEC voltage (VTE)
Reading:
":VTE:MEAS <NR1>"
Select TEC voltage to be measurement value for
“ELCH” on string position <NR1> ( 1....8)
Reading:
":VTE:ACT?"
Reads the actual TEC voltage
[:VTE:ACT
<NR3><LF>]
":VTE:MIN_R?"
Reads U
TE
– ADC = 0000
[:VTE:MIN_R
<NR3><LF>]
":VTE:MAX_R?"
Reads U
TE
– ADC = FFFF
[:VTE:MAX_R
<NR3><LF>]
":VTE:MEAS?"
Reads string position (1....8) of peltier voltage as
measurement value for “ELCH” .
[:VTE:MEAS <NR1><LF>]