Functions in the channel menu
31
_________________________________________________________________________________
_____________________________________________________________________________________________________
© 2010 Thorlabs
The operating mode field and the status field shows the actual status:
CC on/off
constant current mode
CP on/off
constant power mode
AD on/off
sensor is AD 590
TH on/off
sensor is thermistor
Open
TEC output switched on but too high-impedance
ILK
interlock open
OVL
laser diode output open
LIM
laser on but current has reached the limit value
WIN
TEC temperature out of activated temperature window
Vcc
fail
OTP
over temperature
SENS
wrong or no sensor connected
2.2.2 Changing parameters
To set or change a numerical set parameter in the channel menu select the
respective line with the cursor:
Example: I
LD
is to change:
CH1 I
LD
= 44.000mA
CC off P
LD
=0.70000mW
TH off T
A
= +
22.60°C
CHANGE
Internal Power failure, contact
Thorlabs