![THORLABS ITC8 Series Скачать руководство пользователя страница 40](http://html.mh-extra.com/html/thorlabs/itc8-series/itc8-series_operation-manual_1110700040.webp)
34
Functions in the channel menu
_________________________________________________________________________________
______________________________________________________________________________________________________
© 2010 Thorlabs
2.2.7 Selecting the type of the sensor
The sensor type can be chosen in the TED8xxx modules by selecting the line
"Thermistor" respectively "AD590".
AD590
=
AD590 and LM335 families
Thermistor
=
Thermistor. The corresponding range has to be selected
Select the desired type and press
.
2.2.8 Calibrating a thermistor
2.2.8.1 Select the calculation method
If the relation between temperature and resistance for a given thermistor is known,
the PRO8 system is able to display temperature directly in °C instead of resistance in
Ω
. Therefore, a calibration of the sensor in °C is necessary.
Two well known methods to calculate the resistance from temperature are
implemented:
•
The exponential method
•
The Steinhart-Hart method
In the channel menu you can choose between these two methods and enter the
corresponding parameters:
Select the line
Steinh.-Hart
resp.
Exponential
The right softkey toggles between the two methods.
Select the desired type and press
to make setting valid.