Commands
55
_________________________________________________________________________________
_____________________________________________________________________________________________________
© 2010 Thorlabs
3.2.22 Programming the temperature (TEMP)
Programming:
":TEMP:SET <NR3>"
Programs the temperature
":TEMP:MEAS <NR1>"
Programs temperature to be measurement value for
“ELCH” on string position <NR1> (1....8)
Reading:
":TEMP:SET?"
Reads the set temperature
[:TEMP:SET
<NR3><LF>]
":TEMP:ACT?"
Reads the actual temperature
[:TEMP:ACT
<NR3><LF>]
":TEMP:MIN?"
Reads the minimum allowed set temperature
[:TEMP:MIN
<NR3><LF>]
":TEMP:MAX?"
Reads the maximum allowed set temperature
[:TEMP:MAX
<NR3><LF>]
":TEMP:MIN_W?"
Reads T
- DAC = 0000
[:TEMP:MIN_W
<NR3><LF>]
":TEMP:MAX_W?"
Reads T - DAC = FFFF
[:TEMP:MAX_W
<NR3><LF>]
":TEMP:MIN_R?"
Reads T - ADC = 0000
[:TEMP:MIN_R
<NR3><LF>]
":TEMP:MAX_R?"
Reads T - ADC = FFFF
[:TEMP:MAX_R
<NR3><LF>]
":TEMP:MEAS?"
Reads string position (1....8) of temperature as
measurement value for “ELCH”.
[:TEMP:MEAS
<NR1><LF>]
3.2.23 Switching the temperature protection on and off (TP)
Programming:
":TP ON"
Switch temperature protection on
":TP OFF"
Switch temperature protection off
Reading:
":TP?"
Reads status of the temperature protection
[:TP
ON<LF>]
[:TP OFF<LF>]