40
Error messages
_________________________________________________________________________________
______________________________________________________________________________________________________
© 2010 Thorlabs
2. when trying to change parameters during operation:
NOT IF LD ON
It is not possible to change the parameters
LDPOL,
PDPOL, C, Twin
when the laser diode is on
NOT IF TEC ON
It is not possible to change the calibration parameters of
the thermistor, the calibration method or the sensor type
when the TEC is on
3. when trying to switch on LD controller:
OTP
O
ver
T
emperature
P
rotection. Module is overheated and
cannot be switched on.
Vcc fail
Internal supply voltage failure. Please contact Thorlabs!
INTERLOCK
Laser cannot be switched on due to interlock circuit
interruption
Twin
Temperature is out of window (when window is activated)
4. when trying to switch on TEC controller:
NO SENSOR
No sensor connected or wrong sensor found
OTP
O
ver
T
emperature
P
rotection. Module is overheated and
cannot be switched on.
Vcc fail
Internal supply voltage failure. Please contact Thorlabs!
If the error occurs during operation it appears in brackets:
<CH3 interlock>
If the error occurs when switching on or changing parameters it appears in cursor
arrows:
CH3 interlock
Any error must be confirmed by pushing the "
OK
" soft key. Any further operation is
locked until "
OK
" soft key is pushed.