Options
Description
OEM Connection Only
(Administrator only)
Required for API access to the instrument. When enabled, the software (desktop and
Thermo Fisher
™
Connect) cannot connect to the instrument.
API access to the instrument is exclusive to authorized OEM partners.
Thermo Fisher
™
Connect
Region
(Administrator only)
Specify the regional server location to access Thermo Fisher
™
Connect.
Note:
Once you set a region for the Thermo Fisher
™
Connect Platform, restore factory
defaults to change the region.
Date/Time
Set time zone and date and time formats.
Network Connection
Set wired or wireless network connection for the instrument.
Restore Factory Defaults
Restore the instrument to the factory settings.
IMPORTANT!
Back up the instrument before restoring factory defaults (see “Backup
).
Note:
Restoring an instrument to the factory defaults results in the following setting
changes:
·
Instrument profiles and files stored on the instrument are deleted, including all EDS
files and any custom dye calibrations.
·
The most recent valid ROI/Uniformity, Background, and system dye calibrations are
not deleted.
Language
Specify the language to display on the touchscreen.
About Instrument
About Instrument
Displays the instrument
Model Name
,
Block Serial Number
, and
Firmware Version
.
License Agreement
Displays the End User Software License Agreement and the Limited Product Warranty.
You can export the License Agreement to a USB drive.
Notifications
—
Enable home screen notifications of instrument errors and software updates. This
function is not related to the Notifications function in Thermo Fisher
™
Connect.
The number of new, unviewed notifications displays over
Settings
in the home screen.
Maintenance and Service
Software Update
(Administrator only)
Update the instrument software.
Monitoring
(Administrator only)
Enable the
Remote Monitoring Service
to automatically notify Thermo Fisher
Scientific support teams in real-time of potential instrument problems.
The service monitors and sends general
instrument
data, but the service does not
monitor or send
customer
data.
Instrument Statistics
Displays instrument usage information including
Block Cycle Count
and
LED Life
.
Chapter 3
Configure the instrument and manage instrument profiles
Overview of Settings
3
30
QuantStudio
™
1 Plus Real-Time PCR System Installation, Use, and Maintenance Guide