7
Learning Center
Prepare Samples and Blanks
90 NanoDrop Eight User Guide
Thermo Scientific
Run a Blanking Cycle
Run a blanking cycle to verify the following:
• instrument is operating normally (with flat baseline)
• pedestals are clean (i.e., no dried-down sample material on pedestals)
• absorbance contribution of buffer solution you plan to use for sample analyses
Supplies needed
• lint-free laboratory wipes
• calibrated precision pipettor (0–2 µL)
• buffer solution for evaluation
To run a blanking cycle
For quick demonstration, watch multimedia training
Evaluating a Blanking Solution
NOTICE
• Do not use a squirt or spray bottle on or near the instrument as liquids will
flow into the instrument and may cause permanent damage.
• Do not use hydrofluoric acid (HF) on the pedestals. Fluoride ions will
permanently damage the quartz fiber optic cables.
Содержание NanoDrop Eight
Страница 6: ...Contents vi NanoDrop Eight User Guide Thermo Scientific...
Страница 24: ......
Страница 30: ...4 Nucleic Acid Applications 30 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...
Страница 40: ...4 Nucleic Acid Applications Measure Oligo DNA or Oligo RNA 40 NanoDrop Eight User Guide Thermo Scientific...
Страница 56: ...5 Protein Applications Measure Protein A280 56 NanoDrop Eight User Guide Thermo Scientific...
Страница 76: ...6 Custom Applications Measure Custom 76 NanoDrop Eight User Guide Thermo Scientific...
Страница 126: ...7 Learning Center Measurement Screen Display Options 126 NanoDrop Eight User Guide Thermo Scientific...
Страница 142: ...8 Maintenance 142 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...