9
Safety and Operating Precautions
Safety Information
Thermo Scientific
NanoDrop Eight User Guide 149
A UPS reduces the probability of a system shutdown if power is lost elsewhere in the
building. Power line conditioners (which ensure that your service is free from sags,
surges or other line disturbances) also are available in the U.S.A. from us for 120 volt
operation. Line conditioners for 220 volt operation can be purchased locally. Contact
technical support for information about power conditioners and UPS.
Electrical Service Specifications
The following table lists the specifications for electrical service. Contact our service
representative in your area if you have questions about the requirements.
Power Consumption
Generally, 50% more power should be available than the entire system (including
accessories) typically uses. Maximum power consumption and heat dissipation
specifications for the spectrometer and accessories are shown below. The values
are approximate.
Fire Safety and Burn Hazards
To avoid a burn injury and the risk of fire or explosion:
• Use caution when testing flammable or explosive samples (see the “Hazardous
Materials” section)
• Never block any of the vents on the instrument or its power supply
• Only use exact replacement power supplies from us
Requirements
Specifications
Input current
5.0 A (max.)
Input voltage
100-240 VAC
Line frequency
50-60 Hz
Line disturbances
Sags, surges or other line disturbances must
not exceed 10% of input voltage (even for a
half cycle).
Noise
< 2 V (common mode)
< 20 V (normal mode)
Item
Power Consumption
Max. Heat Dissipation
instrument
60 W
205 Btu/hr
Note
Do not position the instrument so that it is difficult to operate the power
switch or access the power supply and power cord.
Содержание NanoDrop Eight
Страница 6: ...Contents vi NanoDrop Eight User Guide Thermo Scientific...
Страница 24: ......
Страница 30: ...4 Nucleic Acid Applications 30 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...
Страница 40: ...4 Nucleic Acid Applications Measure Oligo DNA or Oligo RNA 40 NanoDrop Eight User Guide Thermo Scientific...
Страница 56: ...5 Protein Applications Measure Protein A280 56 NanoDrop Eight User Guide Thermo Scientific...
Страница 76: ...6 Custom Applications Measure Custom 76 NanoDrop Eight User Guide Thermo Scientific...
Страница 126: ...7 Learning Center Measurement Screen Display Options 126 NanoDrop Eight User Guide Thermo Scientific...
Страница 142: ...8 Maintenance 142 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...