Maintenance
Maintaining the Pedestals
132 NanoDrop Eight User Guide
Thermo Scientific
To recondition the pedestals
1. Open the container of PR-1 compound
and use the provided applicator to remove
a pin-head sized amount of the
compound.
2. Apply a thin, even layer of reconditioning
compound to the surface of the upper and
lower pedestal.
Wait 30 seconds for the PR-1 compound
to dry.
3. Fold a clean laboratory wipe into quarters
and use it to vigorously buff the surface of
each pedestal.
Notice
: Support the instrument arm with
one hand while you buff the upper
pedestal to avoid damaging the arm.
Tip
: Black residue on the wipe is normal.
4. Repeat step 3 with a new folded wipe until
all residue is removed and the pedestals
buff clean.
5. Use canned air to remove any paper
residue from the pedestals.
6. Pipette 1 µL DI H
2
O onto the lower
pedestals.
The DI H
2
O should “bead up” or form a
rounded droplet.
Droplet “beads up” on properly
conditioned pedestal
Содержание NanoDrop Eight
Страница 6: ...Contents vi NanoDrop Eight User Guide Thermo Scientific...
Страница 24: ......
Страница 30: ...4 Nucleic Acid Applications 30 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...
Страница 40: ...4 Nucleic Acid Applications Measure Oligo DNA or Oligo RNA 40 NanoDrop Eight User Guide Thermo Scientific...
Страница 56: ...5 Protein Applications Measure Protein A280 56 NanoDrop Eight User Guide Thermo Scientific...
Страница 76: ...6 Custom Applications Measure Custom 76 NanoDrop Eight User Guide Thermo Scientific...
Страница 126: ...7 Learning Center Measurement Screen Display Options 126 NanoDrop Eight User Guide Thermo Scientific...
Страница 142: ...8 Maintenance 142 NanoDrop Eight User Guide Thermo Scientific This page is intentionally blank...