Protein Applications
Measure Protein A280
Thermo Scientific
NanoDrop Eight User Guide 43
Best practices for protein measurements
• Isolate and purify protein samples before measurement to remove impurities.
Depending on the sample, impurities could include DNA, RNA and some buffer
components. See
for more information.
• Ensure the sample absorbance is within the instrument’s
High concentration BSA sample
Low concentration BSA sample
Note
Extraction reagents that contribute absorbance between 200 nm and
280 nm will affect measurement results if present in samples (even residual
amounts).