| Operation
60
|
50155215
Herasafe 2030i / Maxisafe 2030i
Chapter 6
For running the UV disinfection routine, the front window must be completely lowered to
protect against UV radiation. A safety interlock ensures that the routine cannot be run until
the front window is in the closed position.
While the routine is running, the “UV disinfection” screen shown in Figure 46 below is
displayed on the Main Screen.
Figure 46
Main Screen: “UV disinfection” running
The “UV disinfection” screen sports the side-view symbol graphic with the window closed
and a purple orb in the sample chamber. To the right of the graphic a green checkmark
and the “UV disinfection” header identifies the current mode of operation.
Additionally, the “Total cycle time” in hours and the time remaining in hours, minutes and
seconds are displayed.
UV disinfection ends when the preset total cycle time for the routine has elapsed. Then,
the UV lamps are switched off automatically, and the “UV disinfection” screen goes away.
If the setting for “UV disinfection” (see “Setting UV Disinfection Cycle Time” on page 74
for instructions) includes an adjustment for bulb life, the label for the total cycle time
display reads “Total adjusted cycle time”, as shown in the example in Figure 47.
Figure 47
Main Screen: Disinfection running with Adjustment for Bulbs
Tapping the “Total cycle time: n hr” label will take you to the “Disinfection Cycle Settings”
screen if you need to change the setting. The process of setting cycle time is described in
the section “Setting UV Disinfection Cycle Time” on page 74.
You can also tap the “Power Receptacle” button
on or off during an ongoing UV
disinfection cycle.
Содержание HERASAFE 2030i Series
Страница 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Страница 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Страница 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Страница 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Страница 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Страница 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Страница 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Страница 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Страница 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Страница 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Страница 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Страница 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Страница 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...