After the required reaction time, the formaldehyde should be neutralized by evaporating a
25 % ammonium solution (10 ml per cubic meter of sample chamber volume).
Ambient conditions and accessories:
The temperature at the cabinet location should be approx 21 °C, the relative humidity
should be between 60 and 85 %. To evaporate the solution, a heating device with a
container is required.
Cleaning the exterior Surfaces
Wipe the exterior surfaces of the device clean using a solution of tepid water and
commercial dishwasher solution. Then, wipe exterior surfaces dry using a soft, clean cloth.
Cleaning the Front Window
For cleaning the front window (and for disinfecting it using wipe disinfection), the front window
can be closed and hinged.
Cleaning the Drain Pan
Clean the drain pan using a solution of tepid water and commercial dishwasher.
1. Remove the working plates from the workspace.
2. Remove dirt residues and deposits thoroughly.
3. Wipe the drain pan clean using a clean cloth and plenty of clear water.
4. Remove the cleaning liquid from the drain pan and wipe the drain pan surfaces
thoroughly clean.
5. Reinstall the working plates.
NOTE
After the neutralization reaction time with ammonia, carry out check
measurements and ventilate. Ventilation until the MAK value is reached:
Formaldehyde concentration <0.3 ppm (Occupational Exposure Limits).
Material residues
After cleaning, make sure that all cleaning materials have been removed
completely from the drain pan.
Содержание HERASAFE 2030i Series
Страница 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Страница 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Страница 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Страница 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Страница 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Страница 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Страница 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Страница 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Страница 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Страница 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Страница 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Страница 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Страница 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...