Maintenance
Inspection
The annual inspection comprises the following checks:
Electrical safety in accordance with national regulations.
Functional test of the device.
Checking all components for possible damage.
Checking the filter state.
Checking the airflow conditions.
Repeat test in accordance with applicable standard (EN 12469:2000, NSF/ANSI 49
or other.
Service
HEPA filters:
As the filter replacement is an interference with the safety system of the device, filters
must only be replaced by Thermo Electron LED or by adequately trained and authorized
service personnel.
UV Lamps
The UV lamps should be replaced after 8000 operating hours:
1. Remove cover from lamp housing. The cover is only pushed onto the lamp casing
and may be pulled off easily.
2. Rotate the lamp in the socket so that the lamp contacts can be removed from the
groove in the socket.
3. Insert the new lamp into the socket and rotate until the contacts engage.
4. Push the cover onto the lamp casing.
NOTE Perforated plate
The Perforated plate at the blanket of the sample chamber serves for the
protection of the downflow filter and prevents refluxing. While scanning the
filter surface for leak test the perforated plate must be built-in.
NOTE Filter change
Before filters are replaced the disinfection with formaldehyde is mandatory.
After the filter replacement, a repeat test must be performed in accordance
with applicable standard (EN 12469:2000, NSF/ANSI 49 or other).
Before replacing the UV lamps disconnect the unit from the power supply.
Содержание HERASAFE 2030i Series
Страница 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Страница 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Страница 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Страница 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Страница 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Страница 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Страница 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Страница 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Страница 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Страница 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Страница 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Страница 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Страница 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...