| General Notes
8
|
50155215
Herasafe 2030i / Maxisafe 2030i
Chapter 1
Non-intended Use
The safety cabinet must not be used in laboratories that do not comply with the
requirements of safety levels 1, 2, and 3.
The unit must not be operated as a Class II safety cabinet if:
no repeat test is performed after changes to the installation conditions or after
modifications to the technical system.
the alarm system of the device has issued a failure message and the cause for the
failure has not been repaired.
The alarm system must not be tampered with or disabled. If alarm system components
have been removed or disabled for service or repairs, the unit must only be released for
operation if all alarm system components are functioning again properly.
The filters installed in the device are not capable of separating gaseous substances.
Therefore, do not work with or store substances in the device:
are toxic which in quantity or concentration,
if a reaction with other substances may result in hazardous toxic concentrations or
formation of toxic gases,
that may form combustible or explosive mixtures in combination with air.
Содержание HERASAFE 2030i Series
Страница 1: ...Valid 30 08 2017 50155204 21st September 2021 Safety cabinets 50155215 Rev B...
Страница 6: ...Table of Contents 4 50155215 Herasafe 2030i Maxisafe 2030i...
Страница 10: ...Figures 4 50155215 Herasafe 2030i Maxisafe 2030i...
Страница 20: ...Delivery 10 50155215 Herasafe 2030i Maxisafe 2030i Chapter 2...
Страница 130: ...Operation 120 50155215 Herasafe 2030i Maxisafe 2030i Chapter 6...
Страница 134: ...Preparation 124 50155215 Herasafe 2030i Maxisafe 2030i Chapter 7...
Страница 136: ...Shut down 126 50155215 Herasafe 2030i Maxisafe 2030i Chapter 8...
Страница 144: ...Maintenance 134 50155215 Herasafe 2030i Maxisafe 2030i Chapter 10...
Страница 154: ...Technical Data 144 50155215 Herasafe 2030i Maxisafe 2030i Chapter 12...
Страница 156: ...Device Log 146 50155215 Herasafe 2030i Maxisafe 2030i Chapter 13...
Страница 157: ...Herasafe 2030i Maxisafe 2030i 50155215 147 Certificate of Decontamination...
Страница 158: ...Certificate of Decontamination 148 50155215 Herasafe 2030i Maxisafe 2030i Chapter 14...
Страница 173: ...Herasafe 2030i Maxisafe 2030i 50155215 163 Appendix City Time Zones Chapter 15...