![Thermo Finnigan LTQ Скачать руководство пользователя страница 80](http://html.mh-extra.com/html/thermo/finnigan-ltq/finnigan-ltq_hardware-manual_1104223080.webp)
Functional Description
MS Detector
@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@'JOOJHBO-52
5IFSNP
&-&$530/$03103"5*0/
2-26
______________________ Finnigan LTQ Hardware Manual _______________________
Three high-voltage electrical feedthroughs pass through the vacuum
manifold:
•
Two feedthroughs for the high voltage for the conversion dynodes
•
A feedthrough for the RF voltage and DC voltages of the mass analyzer
The vacuum manifold also has an opening for the ion gauge.
The removable top cover plate of the vacuum manifold holds the Q0 and Q1
ion guides, mass analyzer, and two electron multipliers (one part of the ion
detection system). Thus, removal of the top cover plate allows easy access to
these assemblies. Two handles on the top and four guide posts on the
underside of the top cover plate facilitate its removal and installation. An
electrically conductive O-ring provides a vacuum-tight seal between the top
cover plate and the vacuum manifold. The top cover plate and its attached
assemblies are shown in Figure 2-15 and Figure 2-16.
Figure 2-14. Vacuum manifold