MS Detector Maintenance
'JOOJHBO-52
_________________________________________________________ API Source Maintenance
_______________________ Finnigan LTQ Hardware Manual _____________________
4-13
5IFSNP
&-&$530/$03103"5*0/
Removing the Ion Source Interface
To remove the ion source interface, grasp the ridges on either side of the ion
source interface and carefully pull it free from the vacuum manifold. Place it
on a clean surface.
To replace components of the ion source interface, go to the topic: Replacing
Ion Source Interface Components on page 4-15.
To clean the tube lens and skimmer, go on to the next topic: Cleaning the
Tube Lens and Skimmer.
Cleaning the Tube Lens and Skimmer
An accumulation of chemicals on the surfaces of the tube lens and skimmer
forms an insulating layer that can modify the electrical fields that control ion
transmission. The tube lens and skimmer require cleaning less often than the
ion sweep cone and the ion transfer capillary.
To clean the tube lens and skimmer proceed as follows:
CAUTION. Make sure that the LTQ power cord is unplugged before you
proceed.
Figure 4-3. Ion source interface removal
Ion Sweep
Cone
0.3-in. ID O-Ring
Spray Cone
Ion Source
Interface
Vacuum
Manifold
CAUTION. Wait for the ion source interface to cool to ambient temperature
before you disassemble it.