Diagnostics and PCB and Assembly Replacement
Replacing Power Supplies
@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@'JOOJHBO-52
5IFSNP
&-&$530/$03103"5*0/
6-6
_______________________ Finnigan LTQ Hardware Manual _______________________
3. If necessary, remove the left front cover as follows:
a.
If necessary, remove the Ion Max ion source as described in the topic
Removing the Ion Max Ion Source on page 4-12.
b. Unscrew the two fasteners that secure the left front cover to the
vertical beam.
c.
Move the left front cover up by 0.2 in., and then pull it out enough to
access the cables that connect to the back.
d. Disconnect the cables and remove the cover.
4. If necessary, remove the vertical beam.
5. The power supplies in the tower have bulkhead connectors on the back.
To remove a power supply, loosen the fastener that holds it to the chassis
and then carefully tug on the power supply to remove it as a module.
6. Unpack the new power supply. Retain the packing materials so that you
can pack and ship the defective switching power supply assembly to the
Thermo Electron Repair Center in San Jose. Be sure to note the
apparent problem or symptoms on the enclosed forms.
7. Slide the new power supply into the space occupied by the old power
supply. Push carefully on the power supply to seat the connectors in the
rear.
8. Tighten with a Phillips screwdriver the fastener that holds the power
supply to the tower.
9. If necessary, reinstall the vertical beam if you removed it in step 4.
10. If necessary, reinstall the left front cover if you removed it in step 3, as
follows:
a.
While holding the left front cover next to the left side of the MS
detector, reconnect the cables to the back of the cover.
b. Reinstall the left front cover on the MS detector and tighten the two
fasteners that secure the left front cover to the vertical beam.
c.
If necessary, reinstall the Ion Max ion source as described in the topic
Reinstalling the Ion Max Ion Source on page 4-19.
11. Close the right front door of the MS detector. Tighten the Allen screw on
the on the right front door with an Allen wrench.
12. Restart the system as described in the topic Starting Up the System
After a Complete Shutdown in the System Shutdown, Startup, and
Reset chapter.
13. Run the LTQ diagnostics to verify that the system is operational.