Cleaning the Q0 and Q1 Ion Guides
Use the following procedure to remove contamination from the Q0
quadrupole, Q1 octapole, lens L1, and the gate lens. Clean each part in turn.
After cleaning, place each part on a clean, lint free surface.
Figure 4-14. Exploded view of the Q0 and Q1 ion guides
Q1 Octapole
Gate Lens
Mounting Bracket
Thumb Screw
Lens L1
Q0 Quadrupole
Mounting
Bracket
Baffle
Caution. Take care not to bump or jar the Q0 quadrupole and Q1 octapole.
Note. When you clean the ion guide parts, pay particular attention to the
inside surfaces.