MS Detector Maintenance
API Source Maintenance
@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@'JOOJHBO-52
5IFSNP
&-&$530/$03103"5*0/
4-12
______________________ Finnigan LTQ Hardware Manual _______________________
Removing the Ion Max Ion Source
You need to remove the Ion Max ion source to access the ion source interface
and Q00 ion guide.
Remove the Ion Max ion source as follows:
1. Remove the drain tube from the ion source housing drain. See Figure 4-2.
2. Rotate the ion source housing locking levers 90
º
to release the ion source
housing from the ion source mount assembly.
3. Remove the ion source housing by pulling straight off of the ion source
mount assembly, and place the housing in a safe location for temporary
storage.
Go on to the next topic: Removing the Ion Source Interface.
Note. If an ion source probe is still installed in the ion source housing, the
external liquid lines should first be disconnected before removing the ion
source housing.