Functional Description
MS Detector
@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@'JOOJHBO-52
5IFSNP
&-&$530/$03103"5*0/
2-8
_______________________ Finnigan LTQ Hardware Manual _______________________
•
In the Warning condition, the temperature in the MS detector is marginal,
or its +5 V dc is marginal. Marginal temperature is between 37 and 45 °C.
Ma5 V dc supply level is b4.75 and 4.97 V dc. The
Power LED flashes yellow.
•
In the Failure condition, the temperature in the MS detector is greater than
45 °C or its +5 V dc supply level is less than +4.75 V dc. The Power LED
is illuminated solid yellow and the MS detector is held in Reset mode
until the failure condition is cleared.
The LED labeled Vacuum is illuminated green whenever the vacuum
protection circuitry indicates that the vacuum is OK and the safety-interlock
switch on the API source is depressed (that is, the door on the Ion Max ion
source is closed). If the LED labeled Vacuum is not illuminated, high voltage
is not applied to LTQ components.
The LED labeled Communication is illuminated yellow when the MS detector
and the data system are trying to establish a communication link. The
Communication LED is illuminated green when the communication link
between the MS detector and the data system has been made.
The LED labeled System is illuminated yellow whenever the MS detector is in
Standby (that is, high voltage is not supplied to the API source, mass analyzer,
and ion detection system, but the MS detector power is on). The System LED
is illuminated green whenever the MS detector is On (that is, high voltage is
supplied to the API source, mass analyzer, and ion detection system).
The LED labeled Scanning flashes blue whenever the MS detector is On and
scanning ions.
Figure 2-4. Front panel LEDs of the MS detector