MS Detector Maintenance
@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@@'JOOJHBO-52
5IFSNP
&-&$530/$03103"5*0/
4-2
_______________________ Finnigan LTQ Hardware Manual _______________________
Table 4-1.
MS detector maintenance procedures
MS Detector
Component
Procedure
Frequency
Procedure Location
API source
Flush (clean) sample
transfer line, sample
tube, and API probe
Daily
Page 3-6
Ion source interface
Flush (clean) ion
sweep cone and ion
transfer capillary
Daily (or more often
1
)
Page 3-8
Rotary-vane pump
Purge (decontaminate)
oil
Daily
Page 3-11
Ion source interface
Remove and clean ion
transfer capillary
Weekly, or if ion
transfer capillary bore
is contaminated or
obstructed
Page 4-8
PhotoMate light
source
Clean or polish VUV
lamp window
If VUV lamp window is
dirty
Finnigan Ion Max API
Source Hardware
Manual
Ion source interface
Clean tube lens and
skimmer
As needed
1
Page 4-13
Cooling fans
Clean fan filters
Every 4 months
Page 4-44
Q00 ion guide
Clean Q00 quadrupole
and lens L0
As needed
1
Page 4-22
ESI probe
Trim sample tube
If polyimide coating on
the end of the sample
tube has elongated
Page 3-3
APCI or ESI probe
Replace sample tube
If sample tube is
broken or obstructed
Finnigan Ion Max API
Source Hardware
Manual
Ion source interface
Replace ion transfer
capillary
If ion transfer capillary
bore is corroded
Finnigan Ion Max API
Source Hardware
Manual
PhotoMate light
source
Replace VUV lamp
If lamp fails
Finnigan Ion Max API
Source Hardware
Manual
Q0 ion guide
Clean Q0 quadrupole
and intermultiple lens
L1
2
As needed
1
Page 4-28
Q1 ion guide
Clean Q1 octapole and
gate lens
2
As needed
1
Page 4-28