General Description — 37
5. 3. 5. High speed spinning
1. Place the cap
Hold the upper fringe as shown in the figure, and push it onto the shaft downwards.
CAUTION
Make sure that the cap, once set completely, will not come off the shaft. Even when it is pulled holding
the fringe. It is recommended to try setting the cap into the rotor as preliminary practice before spinning
the rotor.
2. Referring to the illustrations, check that the cap is set completely.
➀
The cap is set incompletely.
The cap is set completely.
➀
Push the fringe
3. For the high speed spinning,
see „5. 3. 3. Spinning“ on page 33
5. 3. 6. Deceleration and extraction
1. Regulation of the rotor speed to 3000 rpm by deceleration,
see „5. 3. 3. Spinning“ on page 33
.
2. Push the R. PUMP button and stop the vacuum pump to let air into the rotor chamber.
3. Remove the cap.
Hold the running cap by both hands. Pull it upwards while pushing the center extrusion.
4. Mount the seal assembly.
See “5. 3. 4. Injection of gradient solution and sample” on page 33.
5. Unloading the separated soultion by pump
a. Pipe on the outer wall of rotor
Feed by pump with a heavy solution having a density equivalent to or higher than the maximum density of the
Figure 35
Place the cap
Figure 36
Checking the cap