Thermo Scientific TZ-32 Zonal Rotor
General Description — 32
5. Put the cover onto the rotor containing the septa.
Put the rotor on the fixed holder base and screw in the cover by hand. Mount the handle as shown in the figure to
fasten the cover.
CAUTION
Fasten the cover until the match mark on the cover meets that on the rotor body. As the rotor is used
over and over again, the match mark on the cover will go over that on, the rotor body, which, however,
will result in no trouble.
➀
➁
➂
➀
Handle
➁
Holder base
➂
Match marks
5. 3. 2. Preparation for spinning the rotor
1. Mount the rotor on the spinning shaft in the rotor chamber of the centrifuge.
2. Mount the guard plate.
Secure the fixed piece on the left support first and fix it by turning over the right-hand support toward the right
support.
For setting the supports use the marked upper support hole. Neither mark A nor mark B is employed in the zonal rotor.
Figure 26
Assemble the rotor
cover
Figure 27
Mount the guard
plate