Preventive Maintenance and Servicing
To Replace a Type K Filter Thermocouple
Thermo Scientific
NCAT Asphalt Content Tester
69
10. Test operation of furnace.
To Replace a Type K Filter Thermocouple
1. Disconnect furnace from power supply. (See
for normal cold
resistance value.)
2. Remove the back terminal cover of the furnace. (Note placement and connection of T/C lead
wires.)
3. Remove the screws on the filter thermocouple terminals and disconnect the filter thermocouple
wires.
4. Remove the clamp retaining the thermocouple.
5. Pull the thermocouple straight back.
6. Insert the new thermocouple into the case.
7. Replace the clamp securing the thermocouple.
8. Connect the leads to the terminal block, matching the wire colors with the incoming wires.
9. Replace the back terminal cover on the furnace.
10. Reconnect the furnace to the power supply.
11. Test the operation of the furnace.
To Replace a Solid State Relay
Three solid state relays are located inside the lower control section at the rear. The relays are mounted
on the heat sink inside the case. Facing the front of the furnace, the fan relay is on the right, chamber
elements relay middle and filter relay on the left. See
.
1. Disconnect furnace from power supply.
2. Remove the screws on the front dial plate. Open the front dial plate and remove the two hex 8/32''
(11/32'' socket or wrench is required) nuts in the lower corners behind the front dial plate.
Remove the eight screws on the lower back cover.
3. Remove the (3) four wire modular connectors, one on the right and two on the left, behind the
front cover. (The connectors have a locking tab which has to be opened before removal).
4. Remove the remaining (5) cables from the printer, balance, and display board.
The thermocouple circuit consists of the thermocouple and thermocouple extension
lead wires. Polarity plus to plus and minus to minus must be maintained in the
circuit. Reversing lead wire polarity, thermocouple lead wires touching one another
or grounding to case will cause the chamber temperature to exceed the setpoint
temperature and could cause element failure. The actual displayed temperature will
read on the low side of setpoint temperature.
Содержание Thermolyne NCAT F85930
Страница 6: ...Contents 4 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 10: ...1 Safety Information Warning 8 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 16: ...3 General Specification Environmental Conditions 14 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 18: ...4 Unpacking 16 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 20: ...5 Contents of Carton Accessory Kit Contents Part AY1087X1 18 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 30: ...6 Installation Lift Test Procedures 28 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 36: ...7 Printer Operation Printer Record Storage 34 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 44: ...9 Timer Mode Interrupting the Timer Mode 42 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 50: ...11 Software Features Stability Threshold 48 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 58: ...13 Explanation of Error Codes Setra Balance Error Codes 56 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 65: ...14 Hidden Key Routines Thermo Scientific NCAT Asphalt Content Tester 63 ...
Страница 66: ...14 Hidden Key Routines 64 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 74: ...15 Preventive Maintenance and Servicing To Realign Door Lock 72 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 80: ...17 Exploded View Exploded View Elements 78 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 82: ...18 Ordering Procedures Decontamination Statement 80 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 83: ...Thermo Scientific NCAT Asphalt Content Tester 81 19 Wiring Diagrams LITERATURE CIR WIRE DIAG LT1275X2 ...
Страница 86: ...19 Wiring Diagrams Wiring Diagram Component List 84 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 90: ...20 Amperage Configurations High Amp to Low Amp 88 NCAT Asphalt Content Tester Thermo Scientific ...