6
Installation
Setra Balance Operating Instructions When Used in the Type F85900 NCAT Furnaces
24
NCAT Asphalt Content Tester
Thermo Scientific
connector marked P2. On this same display board, the connector marked P6 should have a red
and white power jack cable plugged into it, and the connector marked P3 should have a 9-pin
ribbon cable connected to it. All three of these connections are necessary in order for the Setra
balance to function.
6. Position the balance so that the pins on the platform support plate are directly beneath the four
tube ports in the bottom of the furnace chamber.
7. Open the furnace door. Insert the four ceramic support tubes through the tube ports located in the
bottom of the furnace chamber. Adjust the position of the balance until the four ceramic support
tubes are seated on the appropriate pins on the balance plate. Finely adjust the position of the
balance until the tubes are centered in their ports, but not touching the sides of the ports.
8. Place the 10" x 11" gray silicon carbide hearth tray on the four ceramic support tubes with the
11" dimension from front to back. Position the front of the hearth tray about one inch back from
the lip of the bottom element plate. Center the hearth tray on the four support tubes, equally
distant from side to side.
9. Close the lower case by returning the control panel to its original position. Secure the control
panel with the four screws you removed in step 1.
10. Switch the main circuit breaker (located on the lower left rear corner of the case when facing the
front of the furnace) on. Close the furnace door and switch the power switch “on”. The balance
display board will read 0.0 grams after approximately 20 seconds of initialization. Turn power
“off” and reposition the balance if the display is unstable or reads anything other than 0.0 grams.
Instability may be caused by one of the support tubes rubbing.
Setra Balance Operating Instructions When Used in the Type
F85900 NCAT Furnaces
ATTENTION OPERATOR!!
Balance Control Ports:ENTER Not used Not used ON/OFF (These ports are used to
calibrate the SETRA balance and change the balance parameters.)
The balance control buttons can be accessed through the pin holes located on the
right front furnace control panel, below the balance indicator, with a tool of .063
diameter or smaller. (See below.)
Содержание Thermolyne NCAT F85930
Страница 6: ...Contents 4 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 10: ...1 Safety Information Warning 8 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 16: ...3 General Specification Environmental Conditions 14 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 18: ...4 Unpacking 16 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 20: ...5 Contents of Carton Accessory Kit Contents Part AY1087X1 18 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 30: ...6 Installation Lift Test Procedures 28 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 36: ...7 Printer Operation Printer Record Storage 34 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 44: ...9 Timer Mode Interrupting the Timer Mode 42 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 50: ...11 Software Features Stability Threshold 48 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 58: ...13 Explanation of Error Codes Setra Balance Error Codes 56 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 65: ...14 Hidden Key Routines Thermo Scientific NCAT Asphalt Content Tester 63 ...
Страница 66: ...14 Hidden Key Routines 64 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 74: ...15 Preventive Maintenance and Servicing To Realign Door Lock 72 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 80: ...17 Exploded View Exploded View Elements 78 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 82: ...18 Ordering Procedures Decontamination Statement 80 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 83: ...Thermo Scientific NCAT Asphalt Content Tester 81 19 Wiring Diagrams LITERATURE CIR WIRE DIAG LT1275X2 ...
Страница 86: ...19 Wiring Diagrams Wiring Diagram Component List 84 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 90: ...20 Amperage Configurations High Amp to Low Amp 88 NCAT Asphalt Content Tester Thermo Scientific ...