14
Hidden Key Routines
Thermo Scientific
NCAT Asphalt Content Tester
59
5 Key-Filter Temperature Adjustments:
6 Key-Disables Error Codes:
7 Key-Displays Actual Filter and Chamber Temperature:
8 Key-Serial Port Output Adjustment:
Access:
Depress the 5 key while simultaneously depressing Power Switch to “ON”.
Procedure:
Adjustment range is 0-900°C. Factory setpoint is 750°C. Chamber
Temperature displays current setpoint. Percent Loss window displays FIL.
Elapsed Time window displays SP for setpoint. Use the keypad to change
the display to the new setpoint value. Depress “ENTER” promptly. This
new setpoint value will be displayed.
Exit:
Depress Power Switch “OFF” and then “ON” to return furnace to normal
operation.
Access:
Depress the 6 key while simultaneously depressing Power Switch to “ON”.
Procedure:
ERR will be displayed in the Chamber Temperature window. “OFF” will
be displayed in the Percent Loss window. The furnace will automatically
return to normal operation within five seconds with the ERR codes
inactivated.
ERR codes are defeated to perform diagnostic procedures as to temporarily
ensure continued furnace operation in the event of a false error code.
Exit:
Depress the Power Switch to “OFF” and then “ON” to return the furnace
to normal operation with activated error codes.
Access:
Depress the 7 key while simultaneously depressing Power Switch to “ON”.
Procedure:
Chamber Temperature will display actual temperature. The Percent Loss
displays actual filter temperature. You can compare actual chamber and
filter temperatures with setpoint to assess the need for the thermocouple or
element replacement or recalibration.
Exit:
Depress the Power Switch to “OFF” and then “ON” to return the furnace
to normal operation.
Access:
Depress the 8 key while simultaneously depressing Power Switch to “ON”.
Procedure:
SPC or SPn will display in the Chamber Temperature window for three
seconds. SPC prints the serial output continuously. SPn prints the serial
output only when the furnace is in the program mode. Accessing toggles
the mode between SPC and SPn. The displayed mode is the active mode.
Default is the SPC mode.
Exit:
Automatically returns to manual operation within three seconds.
Содержание Thermolyne NCAT F85930
Страница 6: ...Contents 4 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 10: ...1 Safety Information Warning 8 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 16: ...3 General Specification Environmental Conditions 14 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 18: ...4 Unpacking 16 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 20: ...5 Contents of Carton Accessory Kit Contents Part AY1087X1 18 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 30: ...6 Installation Lift Test Procedures 28 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 36: ...7 Printer Operation Printer Record Storage 34 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 44: ...9 Timer Mode Interrupting the Timer Mode 42 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 50: ...11 Software Features Stability Threshold 48 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 58: ...13 Explanation of Error Codes Setra Balance Error Codes 56 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 65: ...14 Hidden Key Routines Thermo Scientific NCAT Asphalt Content Tester 63 ...
Страница 66: ...14 Hidden Key Routines 64 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 74: ...15 Preventive Maintenance and Servicing To Realign Door Lock 72 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 80: ...17 Exploded View Exploded View Elements 78 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 82: ...18 Ordering Procedures Decontamination Statement 80 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 83: ...Thermo Scientific NCAT Asphalt Content Tester 81 19 Wiring Diagrams LITERATURE CIR WIRE DIAG LT1275X2 ...
Страница 86: ...19 Wiring Diagrams Wiring Diagram Component List 84 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 90: ...20 Amperage Configurations High Amp to Low Amp 88 NCAT Asphalt Content Tester Thermo Scientific ...