Be sure ambient temperature does not exceed 40°C (104°F). Ambients above this
level may result in damage to or malfunction of the controller. Allow at least 6 inches
of space between the furnace and a surface,and at least 40 inches above the furnace.
This permits the heat from furnace case to escape so as not to create a possible fire
hazard.
Do not install above or near flammables.
6
Installation
Site Selection
20
NCAT Asphalt Content Tester
Thermo Scientific
Site Selection
Place the furnace on a sturdy level surface about 26-28 inches high and capable of supporting the
weight of the fully assembled furnace. Furnace must be placed on a level surface to permit door to
safely swing and remain open for loading and unloading. Allow space for air flow around the furnace
(6" minimum). Locate the furnace so you can direct the furnace’s emissions through ducting to a fume
hood or other appropriate ventilation system. We recommend an exhaust length of not more than
10 feet, if possible. Leveling feet are adjustable. The furnace should be leveled so the door will remain
open.
220-240
20/27
4879/6379
50/60
208
28
5757
50/60
Electrical Connections
The electrical specifications are:
Your furnace is factory configured for a 27 amp supply. You have the option of re-configuring the
wiring at the back of the furnace for 20 amps. Please see
Typically, a 20 amp configuration is used for mobile laboratory installations or if your supply is only
rated for 20 amps.
Prior to connecting your asphalt testing furnace to your electrical supply, be sure the front power
switch and rear circuit breaker are in the OFF position.
The power supply entry is located at the right rear of the lower case (as you face the rear of the unit). A
terminal block is located behind an access plate. Remove the six screws retaining the plate covering the
To avoid electrical shock, this furnace must be installed by a competent, qualified
electrician who ensures compatibility among furnace specifications, power source and
ground code requirements.
For supply connections, use 10 AWG or larger wires suitable for at least 90°C. Failure
to observe this caution could result in damage to furnace and/or work area.
Volts
Amps
Watts
Hertz
Содержание Thermolyne NCAT F85930
Страница 6: ...Contents 4 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 10: ...1 Safety Information Warning 8 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 16: ...3 General Specification Environmental Conditions 14 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 18: ...4 Unpacking 16 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 20: ...5 Contents of Carton Accessory Kit Contents Part AY1087X1 18 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 30: ...6 Installation Lift Test Procedures 28 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 36: ...7 Printer Operation Printer Record Storage 34 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 44: ...9 Timer Mode Interrupting the Timer Mode 42 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 50: ...11 Software Features Stability Threshold 48 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 58: ...13 Explanation of Error Codes Setra Balance Error Codes 56 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 65: ...14 Hidden Key Routines Thermo Scientific NCAT Asphalt Content Tester 63 ...
Страница 66: ...14 Hidden Key Routines 64 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 74: ...15 Preventive Maintenance and Servicing To Realign Door Lock 72 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 80: ...17 Exploded View Exploded View Elements 78 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 82: ...18 Ordering Procedures Decontamination Statement 80 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 83: ...Thermo Scientific NCAT Asphalt Content Tester 81 19 Wiring Diagrams LITERATURE CIR WIRE DIAG LT1275X2 ...
Страница 86: ...19 Wiring Diagrams Wiring Diagram Component List 84 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 90: ...20 Amperage Configurations High Amp to Low Amp 88 NCAT Asphalt Content Tester Thermo Scientific ...