14
Hidden Key Routines
Thermo Scientific
NCAT Asphalt Content Tester
61
Calibration Factor Key-Adjust or Check Test Stability Threshold of a Test:
Temp Key - Toggles from Auto Program Switching Idle to Manual Program
Switching to Idle at Endpoint:
Access:
Depress the Calibration Factor Key while simultaneously depressing the
Power Switch to “ON”.
Procedure:
The stability threshold setting will be displayed in the Percent Loss window.
The Elapsed Time window will display FLN. The value options are 0.01 -
0.5%. Enter the new value via keypad. Observe the Percent Loss window as
the new value is entered. The new value must be entered and the ENTRY
key must be depressed within five seconds.
ASTM Draft 6 specifies test endpoint completion when the measured mass
loss does not exceed 0.01% over three consecutive one minute periods.
AASHTO specifies endpoint completion when the mass loss does not
exceed 0.02% over two consecutive one minute periods. The test stability
threshold value defines the endpoint. The value must match the standard
specification you are applying. The test stability value is printed on the User
Variable Printout. See Hidden Key ENTER. An example of the printout is
depicted on
Exit:
Depress Power Switch to “OFF” and then “ON” to return furnace to
normal operation.
Access:
Depress TEMP key while simultaneously depressing the Power Switch to
“ON”.
Procedure:
AU appears in the Chamber Temp display. The Percent Loss displays either
“ON” or “OFF”. Repeat ACCESS if “OFF” is displayed. Repeating access
changes the display to AU “ON”. You are now in the Auto Program
Switching to Idle mode. Repeat ACCESS again to return to the Manual
Program Switching to Idle mode.
Auto Program Switching to Idle can increase the number of tests run per
day. The operator’s monitoring time is eliminated as the test results
automatically print out at endpoint. The Manual Program Switching to
endpoint is still available in the event continuous running is required after
endpoint.
Exit:
AU “ON” or “OFF” is displayed for approximately five seconds. The
furnace automatically returns to idle.
Содержание Thermolyne NCAT F85930
Страница 6: ...Contents 4 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 10: ...1 Safety Information Warning 8 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 16: ...3 General Specification Environmental Conditions 14 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 18: ...4 Unpacking 16 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 20: ...5 Contents of Carton Accessory Kit Contents Part AY1087X1 18 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 30: ...6 Installation Lift Test Procedures 28 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 36: ...7 Printer Operation Printer Record Storage 34 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 44: ...9 Timer Mode Interrupting the Timer Mode 42 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 50: ...11 Software Features Stability Threshold 48 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 58: ...13 Explanation of Error Codes Setra Balance Error Codes 56 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 65: ...14 Hidden Key Routines Thermo Scientific NCAT Asphalt Content Tester 63 ...
Страница 66: ...14 Hidden Key Routines 64 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 74: ...15 Preventive Maintenance and Servicing To Realign Door Lock 72 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 80: ...17 Exploded View Exploded View Elements 78 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 82: ...18 Ordering Procedures Decontamination Statement 80 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 83: ...Thermo Scientific NCAT Asphalt Content Tester 81 19 Wiring Diagrams LITERATURE CIR WIRE DIAG LT1275X2 ...
Страница 86: ...19 Wiring Diagrams Wiring Diagram Component List 84 NCAT Asphalt Content Tester Thermo Scientific ...
Страница 90: ...20 Amperage Configurations High Amp to Low Amp 88 NCAT Asphalt Content Tester Thermo Scientific ...