4-1
LCD Control Panel
4. LCD Control Panel
This chapter contains details for the centrifuges with the LCD display described in this manual. Pictures shown
are examples and may be different in details to your experience–for example, the LCD display for a ventilated
unit does not have a key for entering temperature, nor a temperature readout.
NOTICE
This chapter shows examples for refrigerated models only.
4. 1. Overview
The LCD display combines a single-line LCD (Liquid Crystal Display) screen with membrane-protected keys for
choosing commands or increasing and decreasing parameter values.
the LCD screen areas and keys described in the following.
➁
➀
➆
➇
➉
➅
➂
➄
➃
➈
➄
No.
Key
Description
➀
Programs
Use the program keys to save and load programs.
➁
Speed
The speed (rpm) or RCF value (x g) is displayed here. You can modify the
value using the
plus
and
minus
key. You can switch between rpm and x g
using the
Arrow
keys.
➂
Time
The running time is displayed here. You can modify the value using the
+
and
-
key.
➃
Temperature
The temperature is displayed here. You can modify the value using the
+
and
-
key. You can pre-temper the centrifugation chamber and the empty rotor
before the centrifugation run starts using the
PreTemp
key.
A
NOTICE
This function is only available on refrigerated centrifuges.
➄
Acceleration /
Deceleration
The acceleration and decelaration profile is displayed here. You can modify
the set profiles using the
+
and
-
key.
➅
Bucket
Use the
Bucket
key in order to have all available bucket types displayed in
succession.
➆
Pulse
Press the
Pulse
key to immediately start the centrifugation run and
accelerate up to the maximal permissible end speed (depending on the used
rotor). Releasing the key initiates a stopping process according to the set
acceleration and braking curve.
➇
Open
Press the
Open
key to activate the automatic door release (possible only if
device is powered on and if the rotor is fully stopped).
➈
Start
Press the
Start
key to start a centrifugation run or to accept the
current settings.
➉
Stop
Press the
Stop
key to manually end the centrifugation run.
Figure 4–1: Functions on the LCD Control Panel
Содержание Sorvall ST Plus Series
Страница 1: ...Thermo Scientific Sorvall X Pro ST Plus Series Centrifuges Instructions for Use 50158527 d 08 2019 ...
Страница 116: ...B 9 TX 750 B 1 4 Biocontainment Certificate ...
Страница 120: ...B 13 TX 1000 B 2 4 Biocontainment Certificate ...
Страница 126: ...B 19 HIGHConic II B 4 4 Biocontainment Certificate ...
Страница 133: ...B 26 Fiberlite F15 6 x 100y B 7 4 Biocontainment Certificate ...
Страница 137: ...B 30 HIGHPlate 6000 B 9 4 Biocontainment Certificate ...
Страница 140: ...B 33 M 20 Microplate B 10 4 Biocontainment Certificate ...
Страница 143: ...B 36 Microliter 48 x 2 B 11 4 Biocontainment Certificate ...
Страница 146: ...B 39 Microliter 30 x 2 B 12 4 Biocontainment Certificate ...